<Preparation of ET Stopper>

Insert the EndoTherapy device into the guide sheath’s 
accessory port so that the device tip protrudes slightly 
from the guide sheath tip. Attach the ET stopper to 
the EndoTherapy device by inserting the EndoTherapy 
device to the end of the slit.

Adjust the position of the ET stopper so that the tip of each device protrudes from the guide sheath tip as 
shown below:

3  Attachment of Stopper to the Guide Sheath

Attach the ET stopper  
to the guide sheath with 
the ultrasonic probe or 
EndoTherapy device 
inside.

1

Insert the guide sheath 
with stopper into the 
endoscope. When 
it reaches the target 
position, move the 
stopper attached to 
the guide sheath until it 
makes contact with the 
biopsy valve.

2

Notes

  · Do not insert the guide sheath into the endoscope when it is not combined with any EndoTherapy accessory or    

 

  ultrasonic probe. Doing so could result in bending or breaking the guide sheath.

 

· When inserting the guide sheath, hold it close to the biopsy valve and keep it as straight as possible, then insert it 

 

  gently into the endoscope.

Tips

Caution

If the endoscope's insertion section near the control 
section is bended, it may be difficult to advance the 
EndoTherapy device into the guide sheath smoothly. 
Make sure to keep the insertion section near the 
control section of the endoscope as straight as 
possible.

Guiding device

Biopsy forceps

Cytology brush

Aspiration needle

Bended

N8602374-122021

Straight

Note:

Make sure the EndoTherapy device 

reaches the end of the slit on the 

stopper.

ET stopper