GuideSheath Kit 2 Quick Reference Guide
Single Use Guide Sheath Kit
K-401/402/403/404
1 Types of Stoppers
Below are important steps for handling the GuideSheath Kit 2. For detailed operating instructions, be sure to
follow the instruction manual included in the device package.
The GuideSheath Kit 2 comes with two types of stoppers – one for an ultrasonic probe and one for EndoTherapy
accessories.
2 Operation of ET/US Stoppers
<Preparation of US Stopper>
Insert the ultrasonic probe into the guide sheath’s
accessory port so that the probe tip protrudes
slightly from the guide sheath tip. Attach the US
stopper to the ultrasonic probe by inserting the
probe to the end of the slit.
Push the US stopper
connection portion
into the guide sheath’s
accessory port to
secure it.
Adjust the position of the
US stopper so that the
probe tip protrudes slightly
from the guide sheath tip.
Remove the US stopper
from the guide sheath’s
accessory port while
attaching the US stopper
to the ultrasonic probe.
1
2
3
Model
K-401
K-402
K-403
K-404
Compatible channel diameter
2.0 mm or more
2.8 mm or more
ET stopper
4 pieces
(White)
3 pieces
(White)
4 pieces
(3: Gray / 1: White)
3 pieces
(2: Gray / 1: White)
US stopper
1 piece
(Gray)
1 piece
(Gray)
Note:
Make sure the ultrasonic probe
reaches the end of the slit on
the stopper.
US stopper
Accessory port