unscrewed, note if there is a small silver wave spring washer adhered to it; if so, remove the
washer from the focus knob and place it on to the brass assembly located within the coarse
focus knob.

Part-2: Installing Flexible Coupling and Baseplate

Slide flexible coupler over threaded end of fine focus shaft while holding right hand knob in place.

Tighten clamp on flexible coupler while holding right hand knob in place.

Slide flexible coupler over the threaded end of the left fine focus shaft while holding the right fine
focus knob in place, and tighten clamp with a 2 mm or 5/64 inch Allen wrench. Ensure that it is
pressed all the way against the brass part of the left microscope knob without pushing the right focus
knob out on the other side. Check for any play in the focus shaft by pushing and pulling the focus
knob as shown on the left – no movement should be evident.  

Part-3: Installing and Aligning the Z-axis Motor Drive

If shipped assembled, use the 7/64 inch Allen wrench and disassemble the Z-drive motor
assembly from base plate by removing the horizontal adjustment screw. At this time, also
loosen the two vertical adjustment screws.

Содержание IX71

Страница 1: ...1 microscope NOTE On January 2007 Olympus changed the size of the threaded end of their ne focus shaft from 3 mm to 4 mm Microscopes with serial numbers 7Axxxxxx and newer and all 8 digit versions ide...

Страница 2: ...The following tools are required for this procedure 1 16 Allen wrench provided 5 64 Allen wrench provided 3 32 Allen wrench provided 7 64 Allen wrench provided The procedure has three parts Removing...

Страница 3: ...the inner left ne focus knob with Allen wrench Loosen set screw in knob if present Use Allen wrench to remove screw that secures the outer focus knob Unscrew inner left ne focus knob while holding the...

Страница 4: ...ding right hand knob in place Slide exible coupler over the threaded end of the left ne focus shaft while holding the right ne focus knob in place and tighten clamp with a 2 mm or 5 64 inch Allen wren...

Страница 5: ...ch Allen wrench to remove the horizontal adjustment screw Use the 7 64 inch Allen wrench to loosen the two vertical adjustment screws Slide the baseplate assembly completely under the microscope as sh...

Страница 6: ...nstall https asiimaging com docs Printed on 2022 01 10 05 45 Slide base plate under microscope and center beneath focus assembly Leave base plate clamp loose at this time Leave base plate clamp slight...

Страница 7: ...in the horizontal adjustment screw to hold the drive assembly in place using the 7 64 inch Allen wrench Leave the adjustment screws loose enough so that later the drive can slide up and down and the...

Страница 8: ...ed or the drive may slip Lightly tighten the vertical and horizontal adjustment screws as shown Since the motor is usually self aligning this should be a good position for the drive Check the alignmen...

Страница 9: ...is felt repeat the above steps Part 4 Installing the Motor Drive Cover Plate and Fine Focus Knob Install cover with screws as shown Screw on small focus knob as shown Locate the motor drive cover Pos...

Страница 10: ...nstall the large outer knob that was removed in Part One Replace the rubber boot if 3 present After screwing on small knob attach the large knob Drive and cover installed on the Olympus IX71 This comp...

Страница 11: ...2022 01 10 05 45 11 11 Olympus IX71 Microscope Z Axis Motor Drive Installation Applied Scienti c Instrumentation https asiimaging com docs...

Отзывы: