Troubleshooting
Inspect the equipment as described in "3 Pre-observation/pre-operation preparation and
inspection" (P. 12), do not use the equipment if there are any obvious malfunctions. Contact
Olympus for repairs. Should even the slightest irregularity be suspected, do not use the instrument.
Perform the actions described in "7.1 Troubleshooting guide" (P. 37). If the problem cannot be
resolved by the described remedial action, stop using the instrument and send it to Olympus for
repair.
7.1
Troubleshooting guide
Error messages
Message
Cause and recommended action
NO RECORDING MEDIA.
An SD card is not loaded.
Load an SD card and try again.
RECORDING MEDIA FULL.
The SD card is full.
Delete unneeded data to free up space or replace with a new SD card.
RECORDING MEDIA IS LOCKED.
PLEASE UNLOCK.
The SD card is locked.
Unlock the SD card you are using and then replace it.
RECORDING MEDIA ERROR.
PLEASE FORMAT THE RECORDING
MEDIA.
The format of the SD card cannot be recognized.
Format the SD card on the instrument.
THIS IMAGE CANNOT BE
RETRIEVED.
The image was not recorded with this instrument.
Only images recorded with this instrument can be displayed.
THE SPECIFIED FOLDER NAME
ALREADY EXISTS.
PLEASE CHOOSE ANOTHER NAME.
A folder with the same name as the one you are creating already exists.
Specify a different folder name.
THE SPECIFIED FILE NAME
ALREADY EXISTS.
PLEASE CHOOSE ANOTHER NAME.
A file with the same name as the one you are creating already exists.
Specify a different file name.
LOAD FILE ERROR.
The free-drawing pattern cannot be loaded.
The data may be faulty. Create the pattern again.
SAVE FILE ERROR.
Free-drawing patterns cannot be created or deleted.
Try creating the pattern again.
IMAGER ERROR.
PLEASE TURN OFF THE POWER.
The self-check function activated and prompted termination of operation.
Stop inspection and turn the instrument off, then on again.
HIGH TEMPERATURE (DISTAL END).
PLEASE IMMEDIATELY PULL OUT
THE INSERTION TUBE.
This message prompts you to stop inspection because the self-check
function was triggered because the distal end was overheating.
Immediately withdraw the insertion tube from the observation object.
NOTE
•
This message appears before the atmospheric temperature of the distal
end of the insertion tube reaches the maximum operating ambient
temperature.