Olympus IPLEX FX Скачать руководство пользователя страница 2

1

6

2

3

4

5

A

 DISTANCE

Distance mode is for measuring the line segment between 

two measurement points

1. Place the cursor on one end of the line segment and press 

MEAS/ENTER on the joystick to specify point A.

2. Place the cursor on the other end of the line segment and press 

MEAS/ENTER on the joystick to specify point B.

l

 The length of line segment A-B is displayed.

Point-to-Line and offset mode are for measuring the distance 

from a measurement point to the reference line defined by 

two points. In Offset mode, an offset line will additionally be 

displayed from the measurement point, in parallel to the 

reference line.

1. Place the cursor on both ends of the line segment and press 

MEAS/ENTER on the joystick to specify points A and B.

l

 Reference line A-B is shown.

2. Place the cursor on the point from which you want to measure the 

distance to reference line A-B and press MEAS/ENTER on the 

joystick to specify point C.

l

 The distance between reference line A-B and point C (or Offset 

Line) is displayed. (Offset mode is only available for IPLEX FX 

models.)

B

 POINT-TO-LINE

C

 OFFSET (FX only)

D

 DEPTH

Depth mode is for measuring the distance from a 

measurement point to the reference plane defined by three 

reference points.

1. Place the cursor on points A, B and C to define the reference 

plane, then press MEAS/ENTER on the joystick to specify points 

A, B and C.

l

 Reference plane ABC is shown.

2. Place the cursor on the point from which you want to measure the 

distance to reference plane ABC and press MEAS/ENTER on the 

joystick to specify measurement point D.

l

 The distance between reference plane ABC and point D is 

displayed. A positive value (height) means the point is closer to 

the tip of the scope than the reference plane, and a negative 

value (depth) means the point is farther away.

E

 AREA/LINES (FX only)

Area mode is for measuring the area defined by multiple 

measurement points. Lines mode is for measuring the total 

length of lines defined by multiple measurement points.

1. 

Place the cursor on points to define the area/lines and press 

MEAS/ENTER on the joystick to specify the measurement points.

l

 

The total length of lines defined by multiple measurement points is 

displayed at the bottom of the screen.

Note: 

Up to 20 measurement points can be specified per measurement 

session.

2. 

When measuring an area, one additional point must be positioned to 

define the complete subject outline. This point (point 6 in the example 

above) needs to be located where the line between this point and the 

last point (point 5) cross across the first drawn line.

l

 

The area enclosed by lines is displayed at the bottom of the screen.

F

 PROFILE (FX only)

Profile mode is for displaying the surface profile of a line 

connecting two points.

1. Place the cursor on both ends of the 

line whose surface profile you want to 

inspect and press MEAS/ENTER on 

the joystick to specify the two points.

l

 The surface profile obtained by 

sectioning a target object with the plane 

defined by the line connecting two 

specified points and the optical center 

is displayed. The surface profile 

graphics overlaid on the right screen 

show the profile viewed from the 

direction perpendicular to the optic 

axis.

G

 MULTI (FX only)

Multi mode automatically detects and displays the 

measurement values of a region specified by two 

reference points. The width, depth and surface area 

of any missing material will be measured.

1. Place the cursor on each side of the material loss and press 

[MEAS/ENTER] on the joystick to specify the reference points. 

The region of missing material will be drawn in the left and right 

screen. The width, depth and area dimension of missing material 

will be displayed at the bottom of screen.

SAVING YOUR STEREO 

MEASUREMENT RESULT

• 

Press and hold the [FRZ/REC] button for at least two seconds to 

save your stereo measurement image with the measurement 

result.

• 

"STILL" indicator will confirm that the image was recorded.

• 

Press [VIEW] button to display the saved image at any time.

Measurement 

point

Measurement 

point

Distance

Measurement 

point

Reference 

point

Perpendicular 

distance

Reference 

line

Offset Line 

(Offset Only)

Reference 

point

Reference 

point

Reference 

point

Measurement 

point

Reference 

surface

Depth or height

A

D

C

B

Reference point

Reference point

Area

Depth

Width

A

B

C

A

B

Measurement 

point

Measurement 

point

Measurement 

point

Measurement 

point