8 |  Attach Suction

While pressing the button on the pinch valve, insert the blue tubing through the slot of the pinch valve and around the strain-
relief hook below. Release the button on the pinch valve. Connect the suction end of the tube set to the vacuum canister.

For best results, use a dedicated vacuum source such as the DIEGO ELITE suction module. If using the DIEGO ELITE 
suction module, insert the filter end of the suction tubing (EGTS102SF) into the suction connection on the front of the 
suction module. Attach the blue suction connector at the other end of the suction tubing to the suction canister.

9 |  Prime

10 | Disassembly

Press and hold down the foot pedal to perform the priming 
sequence. When fluid can be seen flowing out of the distal 
tip of the sheath, the priming sequence is complete.

 Unhook the tube set from the sheath. Slide the sheath distally off the endoscope to disassemble and dispose of the sheath. 
Press the pinch-valve button on the InstaClear console and remove the blue tubing from the pinch valve. Open the pump 
lever and remove the tubing from the pump. Pull the IV spike from the IV bag, being careful not to spill any remaining fluid, 
and dispose of the tube set. Press the power button to turn off the system. Disconnect the power cord from the wall outlet. 
Clean and disinfect the console and foot pedal according to the IFU.