H Y S T E R O F L O W   Q U I C K   R E F E R E N C E   G U I D E

3

C O N N E C T   I N F L O W   T U B I N G

1. Turn the unit on.

2. After self test, press 

Menu

to check

or change system settings. (Refer
to page 4 for menu settings)

3. Hang fluid irrigation bags on the

upper scale hooks.

4. Close inflow and outflow stopcocks

on the instrument.

5. Connect blue Leur Lok on the inflow

tubing to the inflow port on the
 instrument.

6. Press 

OK

next to 

Operative Mode

and press 

Start

.

7. Carefully insert the pressure chamber

on the inflow tubing into the lower
 retainer on the pump, and wrap the
tubing up and around the roller wheel,
locking it into place.

8. Clamp tubing and spike fluid irrigation

bag(s). Open one clamp.

Содержание HysteroFlow/HysteroBalance II

Страница 1: ...HYSTEROFLOW HYSTEROBALANCE II HYSTEROSCOPIC FLUID MANAGEMENT SYSTEM QUICK REFERENCE GUIDE Option 1 Technical Assistance Option 2 Customer Service Option 3 Repair Option 4 Field Inventory...

Страница 2: ...ARNING This quick reference guide is only a summary of the HysteroFlow HysteroBalance II Instructions for Use For detailed operating instructions be sure to follow the Instruction Manual that was incl...

Страница 3: ...ooks 4 Close inflow and outflow stopcocks on the instrument 5 Connect blue Leur Lok on the inflow tubing to the inflow port on the instrument 6 Press OK next to Operative Mode and press Start 7 Carefu...

Страница 4: ...al flow value between 30 and 500 ml min Maximum Deficit Menu Select a deficit threshold value between 200 and 2500 ml Alarm Volume Menu Select a volume between 1 and 5 Bag Size Select a bag size betwe...

Страница 5: ...ow stopcock off when IR is complete Screen prompt and single beep indicate IR completion Do Not Press Stop 1 Place containers on the bottom scale and toggle them with the connection tubes 2 Connect th...

Страница 6: ...ressure Deficit Exceeded 10 Beeps In increments of 100 ml above the preset value Critical Overpressure 10 Beeps Actual Pressure 200 mmHg above nominal pressure Rapid Deficit 10 Beeps Deficit change 30...

Страница 7: ...Outflow tubing set Sterile 5 pkg WA40635A Inflow tubing set Sterile 10 pkg 006775 901 Under Buttocks Drape Fluid Collection Set 5 pkg 006772 901 3L Canisters 36 pkg WA40634A Inflow and Outflow tubing...

Страница 8: ...ot encourage or recommend any actions that deviate from or are not covered by the IFU 2013 Olympus America Inc All rights reserved Trademark and Registered Trademark of Olympus or its affiliated entit...