DMTA034-01EN, Rev. B, March 2014
Chapter 2
30
Figure 2-11 The hexagonal screw that holds the yoke parts together
8.
Ensure that the transducer cable is disconnected, and then unscrew the
transducer from the wedge (see Figure 2-12 on page 30).
It is possible to unscrew the transducer without removing the wedge.
Figure 2-12 Unscrewing the transducer
9.
Screw the new transducer onto the wedge.
Use a hexagonal key
to remove.
Manually turn the
transducer
counterclockwise to
remove; clockwise to
secure.