Log on to with the Username "ALMF".
•
Start the microscope control software using the icon on the desktop
Switching off
•
Check if someone is booked after you on the same day! In case someone is
booked on the same day, just turn the intensity turnknob of the
Argon-laser to the minimum position, switch off the key switch of the
559 nm laser, if there is at least 1 hour to next booking and leave
everything else on.
If nobody else is booked on the same day:
•
Turn the Argon laser key (
d
) to the "Off"-position.
•
Exit the control software.
•
Switch off the HBO lamp.
•
turn the key switch of the 559 nm laser to off and switch off the main
power.
•
Switch off the laser combiner.
•
Switch off the power manager (black main switch).
•
Shutdown the PC.
•
Switch the Argon-laser
main switch off (
c
).