10-003591-01EN [Q7780031], Rev. 4, January 2020
Specifications
57
6. Specifications
6.1
General Specifications
The general specifications for the FlexoFORM scanner are listed in Table 6 on page 57.
Table 6 General specifications
Parameter
Value
General
Dimensions (L × W × H)
26 cm × 10 cm × 10 cm
Weight
1.53 kg
Inspection surface
curvature
From 114 mm (4.5 in.) minimum diameter up to a flat surface
Probe
Type: FA1
64 elements, flexible
Cable length: 5 m
Center frequency: 7.5 MHz
Part number: Q3301201
Wedges
Type: SFA1-Flexo — multiple models for a range of inspection
surface radii (see Table 7 on page 59 and Table 12 on page 71).
Alternative wedges for inspections without a FlexoFORM scanner
are described in “Alternative Components” on page 21, and listed
in Table 13 on page 72 and Table 14 on page 73.
Minimum clearance
Required free access distance over inspection surface:
•
85 mm for 114 mm (4.5 in.) pipe
•
91 mm for 203 mm (8 in.) pipe
•
98 mm for 1219 mm (48 in.) pipe
Содержание FlexoFORM
Страница 6: ...Table of Contents vi 10 003591 01EN Q7780031 Rev 4 January 2020 ...
Страница 8: ...10 003591 01EN Q7780031 Rev 4 January 2020 List of Abbreviations viii ...
Страница 12: ...10 003591 01EN Q7780031 Rev 4 January 2020 Labels and Symbols 4 ...
Страница 24: ...10 003591 01EN Q7780031 Rev 4 January 2020 Important Information Please Read Before Use 16 ...
Страница 26: ...10 003591 01EN Q7780031 Rev 4 January 2020 Introduction 18 ...
Страница 30: ...10 003591 01EN Q7780031 Rev 4 January 2020 Chapter 1 22 ...
Страница 44: ...10 003591 01EN Q7780031 Rev 4 January 2020 Chapter 3 36 ...
Страница 82: ...10 003591 01EN Q7780031 Rev 4 January 2020 Chapter 7 74 ...
Страница 86: ...10 003591 01EN Q7780031 Rev 4 January 2020 Appendix B 78 ...
Страница 90: ...10 003591 01EN Q7780031 Rev 4 January 2020 List of Tables 82 ...
Страница 94: ...10 003591 01EN Q7780031 Rev 4 January 2020 Index 86 ...