Chapter 5 Troubleshooting
67
EVIS EXERA TJF TYPE 160VR OPERATION MANUAL
Distal cover
Others
5.2
Withdrawal of the endoscope with an
abnormality
If an abnormality occurs while the endoscope is in use, take a proper measure
as described in either “When the endoscopic image appears on the monitor” or
“When the endoscopic image does not appear on the monitor or the frozen
image cannot be restored” below. After withdrawal, return the endoscope for
repair as described in Section 5.3, “Returning the endoscope for repair”.
If the endoscope or endo-therapy accessory cannot be
withdrawn from the patient smoothly, do not attempt to
forcibly withdraw it. If any irregularities are suspected,
immediately contact Olympus. Forcibly withdrawing the
endoscope or endo-therapy accessory may cause patient
injury, bleeding and/or perforation.
When the endoscopic image appears on the monitor
1.
Turn OFF all equipment except the video system center, light source and
monitor.
2.
When using the image magnification function of the video system center,
release the function.
Irregularity
description
Possible cause
Solution
The distal cover cannot
be attached.
An improper distal cover
(not MAJ-311) is used.
Use proper distal cover
(MAJ-311).
Irregularity
description
Possible cause
Solution
The remote switch does
not work.
The wrong remote
switch is operated.
Operate the correct remote
switch.
The remote switch
function has been set
improperly.
Set the remote switch function
correctly as described in the video
system center’s instruction
manual.