CONTINUED ON NEXT PAGE
ESG-300
(
Electrosurgical Generator)
Quick Reference Guide
ESG-300 Operation Settings
Basic Procedures
Procedure
Instrument or Technique
Mode
Power
Effect
Snare Polypectomy:
Sessile polyp
Close the snare slowly and gently
PulseCut Slow
120
2
Snare Polypectomy:
Pedunculated polyp
Close the snare slowly and gently
ForcedCoag
20
4
Endoscopic Mucosal
Resection (EMR)
Close the snare slowly and gently
PulseCut Slow
120
2
Hemostasis
Coagrasper
SoftCoag
50
3
Sphincterotomy
Sphincterotome
PulseCut Fast
120
2
Endoscopic Submucosal Dissection (ESD)
Procedure
Instrument or Technique
Mode
Power
Effect
Marking
Knife (apply gentle pressure)
ForcedCoag
20
2
SoftCoag
50
3
Pre-Cut/Incision
Knife
PulseCut Fast
120
2
BlendCut
30
2
Submucosal Dissection
Knife
PowerCoag
30
2
Hemostasis
Closed Knife
PowerCoag
30
2
Peroral Endoscopic Myotomy (POEM)
Procedure
Instrument or Technique
Mode
Power
Effect
Incision
Knife
PulseCut Fast
120
2
Dissection/Tunneling
Knife
SprayCoag
40
2
Myotomy
Knife
PulseCut Fast
120
2
Hemostasis
Coagrasper
SoftCoag
50
3
The above settings are examples based upon previous general clinical usage of the device. Please note that clinical staff are solely responsible for selection of an appropriate
mode and power level, dependent upon the instrument being used and the condition of the tissue being treated. Due to the heterogeneity of instruments and clinical conditions,
Olympus cannot accept liability for clinical risks arising from the use of these example settings.