ESG-300 Operation Settings

Basic Procedures

Procedure

Instrument or Technique

Mode

Power

Effect

Snare Polypectomy: 

Sessile polyp

Close the snare slowly and gently

PulseCut Slow

120

2

Snare Polypectomy: 

Pedunculated polyp

Close the snare slowly and gently

ForcedCoag

20

4

Endoscopic Mucosal 

Resection (EMR)

Close the snare slowly and gently

PulseCut Slow

120

2

Hemostasis

Coagrasper

SoftCoag

50

3

Sphincterotomy

Sphincterotome

PulseCut Fast

120

2

Endoscopic Submucosal Dissection (ESD)

Procedure

Instrument or Technique

Mode

Power

Effect

Marking

Knife (apply gentle pressure)

ForcedCoag

20

2

SoftCoag

50

3

Pre-Cut/Incision

Knife

PulseCut Fast

120

2

BlendCut

30

2

Submucosal Dissection

Knife

PowerCoag

30

2

Hemostasis

Closed Knife

PowerCoag

30

2

Peroral Endoscopic Myotomy (POEM)

Procedure

Instrument or Technique

Mode

Power

Effect

Incision

Knife

PulseCut Fast

120

2

Dissection/Tunneling

Knife

SprayCoag

40

2

Myotomy

Knife

PulseCut Fast

120

2

Hemostasis

Coagrasper

SoftCoag

50

3

The above settings are examples based upon previous general clinical usage of the device. Please note that clinical staff are solely responsible for selection of an appropriate 

mode and power level, dependent upon the instrument being used and the condition of the tissue being treated. Due to the heterogeneity of instruments and clinical conditions, 

Olympus cannot accept liability for clinical risks arising from the use of these example settings.

Содержание ESG-300

Страница 1: ...ase Electrosurgical Generator ESG 300 Bipolar Monopolar On Off Neutral Argon Plasma Unit APU 300 On Off Argon Graphical User Interface Purge button Monopolar Home Screen Active socket Contact Quality...

Страница 2: ...ment Effect level adjustment Open user defined settings System setup Output Socket Name Change Output Sockets Argon Home Screen Foot Switch Indicator Contact Quality Monitor Output Socket Name Argon G...

Страница 3: ...eps 1 2a Turn on the ESG 300 APU 300 Select Mode Monopolar Bipolar Argon Monopolar Bipolar Argon Select mode by tapping on the desired mode located on the bottom of the touch screen 2b Optional Select...

Страница 4: ...bar on the touch screen 2 In the menu screen tap to select the desired mode 3 Once selected the home screen will display the change 4 Changing the Power Level Select to adjust power level 5 Changing...

Страница 5: ...the neutral electrode to the patient 2 Connect to the ESG 300 8 Monitor Neutral Electrodes Contact Quality Monitor If the red neutral electrode lights up alarm tone sounds Is the neutral electrode co...

Страница 6: ...ESG 300 System Operating Steps 9 Bipolar Devices Monopolar Devices Neutral Electrodes Argon Plasma Probes Connect device in appropriate receptacle Creating User Defined Settings STEP 1 On the home sc...

Страница 7: ...Screen OK STEP 1 On the Home screen Select Menu STEP 2 Edit Settings Procedures screen Select New Procedure STEP 3 New Procedures screen Use up down to highlight the setting Use arrow to move the sett...

Страница 8: ...ForcedCoag 20 2 SoftCoag 50 3 Pre Cut Incision Knife PulseCut Fast 120 2 BlendCut 30 2 Submucosal Dissection Knife PowerCoag 30 2 Hemostasis Closed Knife PowerCoag 30 2 Peroral Endoscopic Myotomy POEM...

Страница 9: ...ForcedArgon 50 N A Barrett s Esophagus Activate MAPC probe within a 2 5 mm distance from tissue PulsedArgon Fast 40 N A Pulmonary Procedures Procedure Instrument or Technique Mode Power Effect Less e...

Страница 10: ...on Cutting of varying tissue structures with increased hemostasis capacities Superficial coagulation with increased dissection capability Superficial pinpoint coagulation High peak voltage for superfi...

Отзывы: