Olympus ESG-300 Скачать руководство пользователя страница 1

ESG-300 and APU-300 Display Interface

CAUTION: 

 This guide is only a summary of the instructions for use of the ESG-300 Electrosurgery Generator. Be sure to reference the instruction 

manual that was included with your product purchase.

Electrosurgical Generator  

(ESG-300) 

Bipolar

Monopolar

On/Off

Neutral

Argon Plasma Unit  

(APU-300)

On/Off

Argon

Graphical 

User 

Interface

Purge 

button

Monopolar Home Screen

Active socket

Contact Quality Monitor

Active Mode

Change output socket

Procedure name

Setting name

Power adjustment

Effect level adjustment

Open user defined 

settings

System setup

CONTINUED ON NEXT PAGE

ESG-300

 (

Electrosurgical Generator)

 

Quick Reference Guide

Содержание ESG-300

Страница 1: ...ase Electrosurgical Generator ESG 300 Bipolar Monopolar On Off Neutral Argon Plasma Unit APU 300 On Off Argon Graphical User Interface Purge button Monopolar Home Screen Active socket Contact Quality...

Страница 2: ...ment Effect level adjustment Open user defined settings System setup Output Socket Name Change Output Sockets Argon Home Screen Foot Switch Indicator Contact Quality Monitor Output Socket Name Argon G...

Страница 3: ...eps 1 2a Turn on the ESG 300 APU 300 Select Mode Monopolar Bipolar Argon Monopolar Bipolar Argon Select mode by tapping on the desired mode located on the bottom of the touch screen 2b Optional Select...

Страница 4: ...bar on the touch screen 2 In the menu screen tap to select the desired mode 3 Once selected the home screen will display the change 4 Changing the Power Level Select to adjust power level 5 Changing...

Страница 5: ...the neutral electrode to the patient 2 Connect to the ESG 300 8 Monitor Neutral Electrodes Contact Quality Monitor If the red neutral electrode lights up alarm tone sounds Is the neutral electrode co...

Страница 6: ...ESG 300 System Operating Steps 9 Bipolar Devices Monopolar Devices Neutral Electrodes Argon Plasma Probes Connect device in appropriate receptacle Creating User Defined Settings STEP 1 On the home sc...

Страница 7: ...Screen OK STEP 1 On the Home screen Select Menu STEP 2 Edit Settings Procedures screen Select New Procedure STEP 3 New Procedures screen Use up down to highlight the setting Use arrow to move the sett...

Страница 8: ...ForcedCoag 20 2 SoftCoag 50 3 Pre Cut Incision Knife PulseCut Fast 120 2 BlendCut 30 2 Submucosal Dissection Knife PowerCoag 30 2 Hemostasis Closed Knife PowerCoag 30 2 Peroral Endoscopic Myotomy POEM...

Страница 9: ...ForcedArgon 50 N A Barrett s Esophagus Activate MAPC probe within a 2 5 mm distance from tissue PulsedArgon Fast 40 N A Pulmonary Procedures Procedure Instrument or Technique Mode Power Effect Less e...

Страница 10: ...on Cutting of varying tissue structures with increased hemostasis capacities Superficial coagulation with increased dissection capability Superficial pinpoint coagulation High peak voltage for superfi...