DMTA-10084-01EN, Rev. 2, November 2018
Gates
81
To operate in time-of-flight mode
1.
Open the setup menu, and then select the
Settings
icon (
). See “Setup Mode”
on page 17 for details on using the setup menu.
2.
On the
Settings
menu, select the
Meas Setup
icon (
).
3.
On the
Measurement Setup
page, change the
Units
value to
µs
.
4.
Press the Shift, Tab key sequence to return to the live A-scan display.
In time-of-flight mode, all distance measurements are displayed in microsecond
values instead of inches or millimeters, as indicated above the measurement
reading boxes (see Figure 7-8 on page 81).
Figure 7-8 Time-of-flight measurement
7.8
Zooming in on a Gate
The EPOCH 6LT can quickly zoom in the screen range to provide fine resolution of a
particular inspection zone. When the zoom is activated, the instrument automatically
brings the point that corresponds to the gate start to the left side of the screen and
adjusts the displayed range to match the gate width. The new range equals the normal
gate width. The lowest achievable value of the expanded range is equivalent to the
minimum range of the instrument at the current material-velocity setting. The zoom
flag (
)appears on the right side of the display.
To zoom in on a gate
The zoom function can only be active for one gate at a time.
Measurement units
NOTE
Содержание EPOCH 6LT
Страница 8: ...DMTA 10084 01EN Rev 2 November 2018 Table of Contents viii...
Страница 10: ...DMTA 10084 01EN Rev 2 November 2018 List of Abbreviations x...
Страница 16: ...DMTA 10084 01EN Rev 2 November 2018 Introduction 6...
Страница 20: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 1 10...
Страница 26: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 2 16...
Страница 32: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 3 22...
Страница 68: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 5 58 Figure 5 15 Diagnostic Test setup page...
Страница 109: ...DMTA 10084 01EN Rev 2 November 2018 Calibration 99 Figure 8 7 Calibration Successfully completed...
Страница 110: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 8 100...
Страница 154: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 9 144...
Страница 216: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 11 206 Figure 11 45 Corrosion sidebar menu G2Start...
Страница 220: ...DMTA 10084 01EN Rev 2 November 2018 Appendix 210...
Страница 226: ...DMTA 10084 01EN Rev 2 November 2018 List of Tables 216...
Страница 232: ...DMTA 10084 01EN Rev 2 November 2018 Index 222...