To compensate for the one-way sound path in through transmission (
Thru
)
mode, the
EPOCH 650 does not divide transit time by two when calculating thickness
measurements.
To adjust the test mode
Choose
Pulser > Mode
, and then vary the setting.
5.4.5
Pulser Type
The EPOCH 650 operates in two pulser modes that you can select using the
Pulser
parameter of the
Pulser
mode:
Spike
Used to mimic a traditional spike pulse by using a narrow-width pulse to excite
the transducer.
Square
Used to tune the width of the pulse to optimize the response of the transducer.
The EPOCH 650 uses the PerfectSquare technology to achieve optimum response
from the tunable square wave pulser. This PerfectSquare technology maximizes the
energy used to drive the connected transducer while providing excellent near-surface
resolution.
To adjust the pulser waveform
Select
Pulser > Pulser
, and then vary the setting.
Содержание EPOCH 650
Страница 12: ...DMTA 10055 01EN Rev A February 2015 List of Abbreviations xii...
Страница 30: ...DMTA 10055 01EN Rev A February 2015 Introduction 18...
Страница 54: ...DMTA 10055 01EN Rev A February 2015 Chapter 1 42...
Страница 84: ...DMTA 10055 01EN Rev A February 2015 Chapter 3 72...
Страница 126: ...DMTA 10055 01EN Rev A February 2015 Chapter 5 114...
Страница 148: ...DMTA 10055 01EN Rev A February 2015 Chapter 7 136...
Страница 153: ...DMTA 10055 01EN Rev A February 2015 Programmable Inputs and Outputs 141 Figure 8 1 The A Out setup page...
Страница 154: ...DMTA 10055 01EN Rev A February 2015 Chapter 8 142...
Страница 236: ...DMTA 10055 01EN Rev A February 2015 Chapter 10 224...
Страница 332: ...DMTA 10055 01EN Rev A February 2015 Appendix B 320...
Страница 342: ...DMTA 10055 01EN Rev A February 2015 Appendix C 330...
Страница 346: ...DMTA 10055 01EN Rev A February 2015 Appendix D 334...
Страница 362: ...DMTA 10055 01EN Rev A February 2015 Index 350...