
Important Information - Please Read Before Use
19
OLYMPUS CAPSULE ENDOSCOPE SYSTEM
• Have a spare capsule endoscope ready for use in case the
primary device is determined to be unusable prior to patient
ingestion.
• The recorder unit cradle cannot be used to recharge the
battery pack. To recharge the battery pack, remove it from
the recorder unit, and use the dedicated battery charger.
• After turning ON the capsule endoscope, bring it near the
antenna lead set on the patient, and confirm that the radio
reception indicator lamp blinks in green. If the indicator does
not blink, the capsule endoscope and/or the recorder unit
may be malfunctioning.
• You can check for proper operation of the equipment by
periodically inspecting the status of the indicator lamp.
• If the real time viewer is displaying images passed via the
recorder unit, an icon (
) indicating this will be
displayed on the real time viewer’s display panel.
• When the real time viewer is receiving capsule endoscope
images directly, an icon (
) is displayed on its display
panel.
• The real time viewer is to be used within the hospital only,
and is not to be given to the patient.
Содержание EndoCapsule
Страница 1: ...INSTRUCTIONS OLYMPUS CAPSULE ENDOSCOPE SYSTEM...
Страница 2: ......
Страница 6: ...Contents iv OLYMPUS CAPSULE ENDOSCOPE SYSTEM...
Страница 8: ...2 Labels and Symbols OLYMPUS CAPSULE ENDOSCOPE SYSTEM Do not iron Do not dry clean...
Страница 34: ...28 Important Information Please Read Before Use OLYMPUS CAPSULE ENDOSCOPE SYSTEM...