DMTA-20006-01EN, Rev. B, February 2014
Setup and Operation
23
Figure 2-2 Installing the protective sleeve
3.
Install the sleeve clamp as follows (see Figure 2-3 on page 23):
a)
Install the sleeve clamp over the bundle and then slide it over the protective
sleeve.
b)
Install the sleeve clamp screw into the sleeve clamp so the threads of the
sleeve clamp screw just start to engage in the sleeve clamp (see illustration A).
c)
Slide the protective sleeve over the cable-management portion of the tail.
d)
Tighten the sleeve clamp screw until sufficiently snug (see illustration B).
e)
Ensure that the sleeve clamp is oriented so that the sleeve clamp screw does
not protrude lower than the tail and rub on the inspected surface.
Figure 2-3 Installing the sleeve clamp
4.
Assemble the probe and the wedge:
a)
Select the appropriate wedge for the outside diameter of the pipe to be
inspected (see Table 15 on page 65 and Table 16 on page 66).
b)
Apply couplant on both the probe and the wedge-surface interface.
c)
Screw the probe on the wedge.
A
B
C
D
A
B