Olympus CKX41 Скачать руководство пользователя страница 9

Part

1.  Inspection items and Methods

<CKX31>

Item

Standard

Method

Binocular

tube

Interpupillary
distance adjustment

48mm - 75mm

In the observation state, insert a thin sheet of a

paper with graduations at the eye point position and

 

measure the interpupillary distance.

Interpupillary
distance working force

(500g - 1500gf)

5N - 15N

0.1mm or less on the image

surface

Put a string on the sleeve and measure the working

 

force to actuate the interpupillary distance using

 a tension gauge.

Using the standard eyepiece (KN0048; with

 adapter-1) and an objective (PlanC10X),

 

observe a specimen whose center can be identified

 (ex. concentric circles) on the stage.

By moving the stage, match the specimen center
and the visual field center in the right sleeve.

range

Revoving axis

 

Open and close the interpupillary distance, and

read the movement of the image using  the reticle

scale (1 graduation = 0.1mm) of KN0048.

Left / right optical axis

On image surface:

0.2mm or less in vertical
direction
0.2mm or less in outward
direction

0.4mm or less in inward

direction

Using the standard eyepiece (KN0048; with

 

adapter-1) and an objective (PlanC10X),

 

observe a specimen whose center can be identified

 (ex. concentric circles) on the stage.

Taking the right sleeve as the reference, read the

displacement between the specimen center and

the visual field center in the left sleeve using

the reticle scale (1 graduation = 0.1mm) of the
KN0048.

Absolute optical axis

0.1mm or less on the image

surface

Combining the standard eyepiece (KN0048; with

Adapter-1), a microscope frame (product), the

 standard objective (KN0041), read the

displacement between specimen center in KN0041

and the visual field center in the right sleeve
using the reticle scale (1 graduation = 0.1mm) of

the KN0048.

Exit pupil center

Within 30% of objective’s

exit pupil diameter

Combining the centering telescope (KN0029),
an objective (PlanC10X), and microscope frame

(product), read the displacement between the exit

pupil center of the objective and cross hairs center

Pafocality

+ / - 0.3mm

Combine the standard eyepiece (KN0048; with

adapter-1, the focusing telescope (FT-36),

the microscope frame (Product), the standard

objective (KN0041).

 In the left sleeve, focus on the specimen in KN0041
 and check the parfocality using the helicoid scale

 

(1 graduation = 0.1mm) of the KN0048.

* 1) Since the  binocular section of CKX31 is the same as that of CK30, the procedure of optical axis adjustment is

 

performed in the same manner as CK30. (Refer to the adjustment part of D-3 - D-14  in CK30/40 repair manual.)

 

 2) Nesessary jigs are partially different because UIS optical system is adopted in CKX. (Refer to the above description.)

  3) Diffrence between CK30 and CKX31 in parfocality adjustment: CK30  uses the spacer (selection : 3pcs. available),

 

 While CKX31 uses the washer ( selection: 6 pcs. available) when adjusted and it requires more precise standard.

of centering telescope (KN0029).

B-1

CKX31/CKX41

                  

   B.  INSPECTION STANDARD

(center of standard

objective)

 

Tension gauge (30N): OT3223

Содержание CKX41

Страница 1: ...CULTURE MICROSCOPES CKX31 CKX41 REPAIR MANUAL SCIENTIFIC EQUIPMENT DIVISION MARKETING DEPARTMENT All rights reserved Reproduction in whole or in part without written permission is prohibited 2H ...

Страница 2: ...anuals because the explanation of these parts are ommited as a common repair skill In optical adjustment section parfocality adjustmet is mainly explained on D 6 and D 12 The parfocality is required to adjust precisely due to severe standard compare with CK30 40 2 Before optical adjustment refer to jigs used in CKX31 CKX41on D 4 D 9 and D 10 5 Since the descriptions of disassembling and assembling...

Страница 3: ...rical unit D 14 F LUBRICANTS AND CHEMICALS 1 Listofgreases F 1 E JIGS AND TOOLS 2 List of adhesives F 1 D REPAIR PROCEDURE 1 Overview of optical adjustment D 1 2 Adjustment procedure CKX31 D 3 2 1 Preparation for optical adjustment CKX31 D 4 2 2 Checking parfocality for CKX31 D 5 2 3 Parfocality adjustment for CKX31 D 6 3 Adjustment procedure CKX41 D 8 3 1 Preparation for optical adjustment CKX41 ...

Страница 4: ... changing the objective when combined with IX2 SLP Applicable when the CKX31 CKX41 is combined with the IX2 SLP 3 Main switch and light intensity adjustment knob are located on the front of microscope frame Slider Magnification 10X 20X 4X 40X IX2 SLP dedicated RS built in PHL built in RS adjustment IX2 SL Common to 10X 40X built in Unnecessary IX2 SLPH1 IX2 SLPHC IX2 SLPH2 RS adjustment unnecessar...

Страница 5: ...ntrast effectiveness can not be obtained In a combination of IX2 SL objective other than the above appicable objective is possible to use as long as the objective corresponds to PHL PHC PH1 PH2 2 Brightfield observation PLC4X PLC10X ACH20X LCPLFL40X Since illumination does not cover the N A of LCPLFL40X objective resolution is not In brightfield observation objective other than the above ones can ...

Страница 6: ...der Substage 70 L X180 W mm Auto X Y stage spot scan mountable 4 5 Illumination system Light source 6V30W halogen bulb 6V30WHAL Average life time of bulb 100 hours Lamp socket U LS30 3 2 Filter holder Detachable filter holder 45mm filters can be inserted up to 11mm in thickness Aperture diaphragm lever control type diameter 3mm 44mm minimum maximum Slider insertion With phase slider pocket Built i...

Страница 7: ...5 Dimensions 364 370 5 486 6 194 5 380 5 473 2 364 370 5 515 5 195 393 1 473 7 1 CKX31 1 CKX41 UNIT mm 237 6 237 6 CKX31 CKX41 A OUTLINE OF PRODUCT A 4 ...

Страница 8: ...1 LCACH40X2PH PH2 CPL10XRC LCACH20XRC LCACH40XRC 45KB 3 L U RFLT50 32ND25 CKX RFA w B G WH10X WH10X H 35WH10X PWH10X U CT30 WHB10X CT 5 WHB10X H CK40 CPG CK40 CPG30 IX2 GCP CKX41SF w U LS30 3 2 CKX31SF New unit primary New unit second LUCPLFL40XRC Others objectives applicable U LH50HG CKX NU w U LS30 3 2 6V30WHAL 6V30WHAL In U excitation remove the UV cut filter For U excitation w WHB10X WHB10X H ...

Страница 9: ...ion 0 1mm of the KN0048 Absolute optical axis 0 1mm or less on the image surface Combining the standard eyepiece KN0048 with Adapter 1 a microscope frame product the standard objective KN0041 read the displacement between specimen center in KN0041 and the visual field center in the right sleeve using the reticle scale 1 graduation 0 1mm of the KN0048 Exit pupil center Within 30 of objective s exit...

Страница 10: ...kness Thickness gauge OT1949 Part Item Standard Method Exit pupil center Within 10 of objective s exit pupil diameter Combining the centering telescope KN0023 an objective PlanC10X UIS observation tube BXKN001 and microscope frame product read the displacement between the exit pupil center of the objective and cross hairs center of centering telescope KN0023 Pafocality 0 3mm Combine the standard e...

Страница 11: ... Refer to the above description method exit pupil center and check the optical axis and exit pupil center 1 Revolving axis 1 Standard eyepiece KN0048 with adapter 1 2 Objective PlanC10x 3 Test plate concentric circles KN0003 or cross micrometer OB M1 100SQ referring the CH3 BI45 repair manual 2 Left right optical axis Same as the above 3 Absolute optical axis 1 Standard eyepiece KN0048 with adapte...

Страница 12: ...Screw mounting part Inner screw threads Grease OT2144 Inner screw threads Grease OT2144 Screw it to G Change the position to adjust absolute optical axis Parfocality adjustnent 1 CKX31 frame b Selection CKX31 CKX41 C DISASSEMBLY AND ASSEMBLY PROCEDURES C 1 ...

Страница 13: ...fer to C 15 ACU3X6SA 1pc 7 OT2144 Adjust absolute optical axis For parfocality adjustment refer to D 6 Relay lens part when adjusting parfocality remove the lens frame 1 a and insert the selected washer under it If necessary remove the lens frame 3 c from the frame 2 d Washer selection AE092300 t 1mm standard AE092100 t 0 5mm AE092200 t 0 7mm AB129900 t 0 05mm AB130000 t 0 1mm AB130200 t 0 3mm CKX...

Страница 14: ...rame A B 1 D 2 C E 3 4 F G 5 6 H I a b c d 7 e 8 f g Screw mounting part Adhesive OT1026 Inner screw threads Grease OT2144 Screwit to H Disassembly of Mount g C 3 CKX31 CKX41 C DISASSEMBLYANDASSEMBLY PROCEDURES ...

Страница 15: ...tical axis For parfocality adjustment refer Relay lens part I when adjusting parfocality remove the lens frame 1 a and insert the selected washer under it If necessary remove the lens frame 3 c from the lens frame 2 d Washer selection AE092200 t 0 7mm standard AE092100 t 0 5mm AE092300 t 1mm AB129900 t 0 05mm AB130000 t 0 1mm AB130200 t 0 3mm CAP COVER COVER CK4 ILL30 STAGE PART CAP LENS FRAME ASS...

Страница 16: ...Grease OT3189 Grease OT3189 Grease OT3189 Contacting surface Grease OT3189 Tension adjustment heavy light Prism ass y pushing direction Prism ass y pushing direction only Adhesive tape C 5 CKX31 CKX41 C DISASSEMBLY AND ASSEMBLY PROCEDURES screw for fixing eyepiece screw for fixing eyepiece 6 Double sided ...

Страница 17: ...onvex of sleeve ass y G HELICOIDASS Y F CUK3X6SB 2pcs 7 Refer to D 1 Revolving axis Follow the adjustment procedure on D 3 Left right optical axis Follow the adjustment procedure on D 3 Push it in the arrow directions during assembling Exit pupil center Follow the adjustment procedure on D 3 Same as the above Apply grease to the contacting surface only Absolute optical axis Follow the adjustment p...

Страница 18: ...Tension adjustment Adhesive tape Grease OT3189 Grease OT3189 Grease OT3189 Grease OT3189 Scale seal Contacting surface Prism ass y and inclined prism Pushing direction when assembling Making it even in left and right direction 11 C 7 CKX31 CKX41 C DISASSEMBLY AND ASSEMBLY PROCEDURES 6 6 When removing diopter ring F Loosen the scews 5 and remove the screw 6 for fixing eyepiece Double sided ...

Страница 19: ... Same as the above Apply grease to the contacting surface only Absolute optical axis Follow the adjustment procedure on D 8 CUTB3X10SA 2pcs 3 BINOCULAR TUBE CUKK3X6SA 4pcs 4 I T U V FIXING PLATE PRISM CUK3X6SA 2pcs 12 F DIOPTER RING ACU3X3SA 2pc 5 Since the mounting procedure is the samet as that of CKX31 K TORQUE PLATE AWU3X4SA 2pcs 11 L SPRING WASHER Since the adjustment is the same refer to C 6...

Страница 20: ...ing guide unit M when assembling it push it in arrow directions Grease OT2008 Grease OT2008 Adhesive tape 8 Adhesive OT1026 Grease OT2006 Grease OT2006 Grease OT2012 Grease OT2012 Grease OT2012 gears f g h Grease OT2008 Adhesive OT1378 screw threads Adhesive OT3111 screw threads Check item refer to B inspection standard on B 2 4 pcs 10 4 pcs C 9 CKX31 CKX41 C DISASSEMBLY AND ASSEMBLY PROCEDURES 5 ...

Страница 21: ...ce of the shaft a Apply grease to the outer surface Apply grease to the outer surface While holding the coarse adj knob ass y screw the fine shaft mount into it Check that the gears moves smoothly Apply grease to the outer surface Rotating tension Standard 0 3N 1 2N OT3225 Rotating tension Standard 6N 25N OT3223 Apply grease to the shaft hole Fit the protrusions into the holes of frame Push it in ...

Страница 22: ...uide M Arrange the rollers altenately on the upper side so that they face in same direction to lower side Adjustment guide mounting position Working force standard 0 2 0 5N without play or tightness Guide D Grease OT2010 Apply grease to sliding surface of wire guide Fit the protrusions into the holes of frame Adhesive tape Rack Adhesive OT1838 CKX31 CKX41 C DISASSEMBLYANDASSEMBLYPROCEDURES 6 Focus...

Страница 23: ... without play or OT3225 tightness Check the sliding surface of wire guide If there is a dent on it replace the wire guide Grease applied part each groove attached wire guide sliding surface of wire guide E Apply grease to rollers Apply adhesive to fixing screws in the arrow directions when replaced Refer to the figure on C 11 Apply grease to rollers OT2010 OT2010 Arrange the rollers alternately OT...

Страница 24: ... 11 a screws for fixing the bottom plate Also remove the screws 2 b screws for fixing the circuit board 8 9 10 3 Rheostat ass y CN1 to CN2 CN3 CKX31 CKX41 C DISASSEMBLYANDASSEMBLY PROCEDURES C 13 7 Electrical unit VR21 VR23 Voltage adjustment trimmers VR21 minimum voltage VR23 maximum voltage Refer to D 14 ...

Страница 25: ...ector Connection to CN3 rheostat ass y connector Connection to CN2 lamp connector Voltage adjustment it is necessary to adjust it if the rheostat ass y or circuit board F is replaced individually Refer to D 14 It is used to fix the cover I to the base of frame CUK3X6SA 4pcs 6 SW3SA 4pcs 7 CUK3X6SA 1pc 8 SW3SA 1pc 9 KNW3SA 1pc 10 CUKK3X5SA 2pcs 11 Be sure to turn off power during disassembly and re...

Страница 26: ... 5 6 W X Y CKX31 CKX41 C DISASSEMBLYANDASSEMBLY PROCEDURES C 15 8 Transmitted illuminator Adhesive tape Adhesive OT1026 Adhesive OT1026 Adhesive OT1026 Adhesive OT1026 Grease OT2010 Grease OT2010 slidingsurface on the bottom sliding surface on the top Double sided Double sided ...

Страница 27: ...ER Attach it with double coated adhesive tape G H CUKSK3X8SA 2pcs 3 L CSK3X6SA 3pcs 4 PSK2X4SA 3pcs 5 OT1026 N OT1026 Apply adhesive to 2 points V RING SCREW SPRING BALL X Y Z OT1026 Mount it in the correct directions Refer to the figure on C 15 AB4X10SA 1pc 2 CKX31 CKX41 C DISASSEMBLYANDASSEMBLYPROCEDURES C 16 ABK2 6X8SA 4pcs 6 OT2010 OT2010 Attach it with double coated adhesive tape on the botto...

Страница 28: ...parfocality in the left sleeve After checking it in the left sleeve turn the helioid to focus on the specimen in KN0041 then loosen the screws 1 only move the diopter ring and fix it at the position where moving range from focusing point is equal in scewing in out directions 1 50 60 70 75 1 After removing the stage peel off the seals and remove 1 Screws 3PUTB2X4SA 1pc 1 CUKSK3X8SA 4pcs 2 seal Cove...

Страница 29: ...2 Cover Cover L Cover R Fig 2 Fig 2 4 Parfocality CKX41 In the same manner as CKX31 adjust the paforcality by inserting washer under the lens frame 1 For adjustment procedure refer to D 12 2 1 Left sleeve After checking it in the right sleeve turn the helioid to focus on the specimen in KN0041 then loosen the screws 1 only move the scale ring and fix it at the position where scale 0 is aligned to ...

Страница 30: ...0 40 repair manual Jigs See D 4 Procedure See D 8 4 2 4 3 to D 10 in CK30 40 repair manual Jigs See D 4 Procedure See D 11 5 2 to D 13 in CK30 40 repair manual Jigs See D 4 Procedure See D 14 6 2 in CK30 40 repair manual See D 5 to D 6 See D 17 in CK30 40 repair manual See C 5 to C 6 In adjustment of revoving axis and left right optical axis it is nesessary to use CK40 MVR and supplied slide glass...

Страница 31: ...m OT1949 KN0048 Adapter 1 2 PlanC10X 3 Absolute optical axis 1 Standard eyepiece KN0048 Adapter 1 2 Standard objective KN0041 2 PlanC10X 5 Parfocality 1 Standard eyepiece KN0048 Adapter 1 3 Standard objective KN0041 2 Focusing telescope FT 36 or U FT 2 1 Preparation for optical adjustment CKX31 KN0003 or cross micrometer OB M1 100SQ 3 Test plate concentric circles KN0003 or OB M1 100SQ cross micro...

Страница 32: ...it does not meet the standard adjust the parfocality referring to D 6 Standard 0 3mm within 3 graduations 2 2 Checking parfocality for CKX31 Necessary jigs 2 Standard eyepiece KN0048 Adapter 1 1 Focusing telescope FT 36 or U FT Set the following jigs as illustrated below to check the parfocality 3 Standard objective KN0041 Insert it into left sleeve CKX31 CKX41 D REPAIR PROCEDURE D 5 ...

Страница 33: ... selection 5 Assemble the parts in reverse order of disassembly 6 Set the jigs focus on the reticle of KN0048 and the specimen in KN0041 At this time read the difference from the index on the helicoid scale Standard 1 Remove the stage part and take off the covers refer to the disassembly on C 1 A B C D Relay lens Ass y BI tube Relay lens ass y consists of three parts lens frame 1 A frame 2 B and l...

Страница 34: ... is 1 9 mm when the total thickness 1 70mm of washers 0 7mm 1pc and 1mm 1 pc are inserted parfocality difference becomes 0 06mm reading value standard 1 0 Unit mm 0 05 0 1 0 3 0 5 0 7 1 0 1 1 1 70 0 06 1 70 1 65 0 04 0 00 1 60 0 03 1 55 0 06 1 1 1 1 55 0 04 1 50 0 01 1 45 0 02 1 40 0 05 1 1 1 1 1 40 0 05 1 35 0 02 1 30 0 01 1 1 1 25 0 04 1 25 0 06 1 20 0 02 1 15 0 01 1 10 0 04 1 10 0 06 1 05 0 03 ...

Страница 35: ... D 10 Procedure See E 8 to E 9 in CH3 BI45 repair manual Jigs See D 9 and D 10 Procedure See E 10 to E 11 in CH3 BI45 repair manual Jigs See D 9 and D 10 Procedure See D 14 6 2 in CK30 40 repair manual See D 11 to D 12 See D 17 in CK30 40 repair manual See C 6 to C 7 In adjustment of revoving axis and left right optical axis it is nesessary to use CK40 MVR and supplied slide glass holder when movi...

Страница 36: ...r 2 2 3mm adapter ring KC2040 3 Focusing telescope FT 36 or U FT 4 Standard objective KN0041 5 UIS standard observation tube BXKN001 4 Stage squareness 1 Stage for checking stage tilt alignment SKN0003 2 Thickness gauge 0 3mm OT1949 KN0048 with adapter 1 KN0023 Adapter 2 KC2040 BXKN001 KN0041 PlanC10X SKN0003 3 1 Preparation for optical adjustment CKX41 FT 36 CKX31 CKX41 D REPAIR PROCEDURE For opt...

Страница 37: ...cality 1 Standard eyepiece KN0048 Adapter 1 3 Standard objective KN0041 2 Focusing telescope FT 36 or U FT KN0003 or cross micrometer OB M1 100SQ 3 Test plate concentric circles KN0003 or OB M1 100SQ cross micrometer right sleeve left sleeve right sleeve Optical adjustment in binocular tube KN0003 D 10 KN0041 PlanC10X Parfocality Check the parfocality in right sleeve If not meet the standard confi...

Страница 38: ...1 Necessary jigs 2 Standard eyepiece KN0048 Adapter 1 and Adapter 2 5 Standard objective KN0041 1 Focusing telescope FT 36 or U FT Set the following jigs as illustrated below to check the parfocality 3 3mm ring adapter KC2040 4 UIS standard observation tube BXKN001 KN0041 If it does not meet the standard adjust the parfocality by inserting washer refer to D 12 Standard 0 3mm within 3 graduations D...

Страница 39: ...d turn the relay lens ass y counterclockwise to remove it 2 1 2 Remove the screws 1 and remove the part of relay lens ass y and TR body ass y Screw AB3X14SA 1 Washer BNW3SA 4pcs each Screw ACU3X6SA 2 1pc 4 Remove the lens frame 1 ass y A and insert appropriate washer selection 5 Assemble the parts in reverse order of disassembly 6 Set the jigs focus on the reticle of KN0048 and the specimen in KN0...

Страница 40: ...differnce is 2 0 mm when the total thickness 1 05mm of washers 0 05mm 1pc and 1mm 1 pc are inserted parfocality difference becomes 0 03mm reading value standard 0 7 Unit mm Numeric character 1 or 2 in the column of washer selection shows the number of washers 0 05 0 1 0 3 0 5 0 7 1 0 1 1 1 1 05 0 13 1 1 00 1 00 0 06 0 04 1 1 1 00 0 14 0 95 0 05 1 2 1 0 95 0 05 0 90 0 14 0 90 0 04 2 1 0 90 0 06 1 1...

Страница 41: ...bly procedures on C 13 to C 14 A A Power cord VR21 VR22 VR23 6V30W halogen bulb Lamp cable a Minimum voltage adjustment 1 Turn ON the power Turn the light Power switch Set the digital multimeter to CN2 2 pin CN2 intensity control part a cunterclock wise to lower lamp brightness to its lowest level 2 Rotate the circuit board s trimmer VR21 between the CN2 s 1 and 2 pins is within to adjust so that ...

Страница 42: ...ntering telescope Standard objective Standard eyepiece including adapter 1 and 2 Thickness gauge Tension gauge 30N Tension gauge 3N Gauge for checking stage tilt alignment Test plate 5 100mm concentric circles B 2 D 9 11 12 B 1 2 3 D 4 5 9 10 11 B 2 D 9 11 B 1 3 D 4 10 B 2 D 9 B 1 3 D 4 10 B 1 2 3 D 4 5 6 9 10 11 12 B 2 D 4 9 B 1 2 C 6 8 10 B 2 C 10 12 B 2 D 4 9 Digital multimeter D 14 or OB M1 10...

Страница 43: ...light Silicone grease Grease medium Grease heavy Mo grease 2 List of adhesives No Description Page OT3111 OT1378 OT1026 OT1838 Cyanoacrylate adhesive Solvent based adhesive transparent Solvent based adhesive red Cyanoacrylate adhesive F 1 C 2 4 C 6 8 C 10 C 10 C 10 C 2 4 10 16 C 10 C 12 C 10 OT2010 Grease light C 12 16 ...

Отзывы: