Страница 1: ...BINOCULAR TUBE _I CH3 Bl45 REPAIR MANUAL OLYMPUS 5 All rights reserved Reproduction in whole or in part without written permission is prohibited 61 01 10 1997...

Страница 2: ...3 4 Optical Adjustment Mechanism i I i B 4 INSPECTION STANDARD 1 CheckItems andMethods C l 3 Useof Jigs andTools c 2 DISASSEMBLY AND ASSEMBLY PROCEDURES 1 Entire DisassemblyandAssembly D l 2 Division...

Страница 3: ...f performance r f 3 It is possible to observe in wide range of field number 20 Conventional field number was 18 4 The interpupillary distance is adjustable from 48 mm to 75 mm applicable to various ob...

Страница 4: ...5 mm adjustmendrange 5 Magnification lx 6 Image direction Inverted image 7 Left right diopter difference Helicoid ring for adjustmentprovided in compensation the left sleeve Diopter range 2 5 diopters...

Страница 5: ...CH3 Bl45 A OUTLINE OF PRODUCT f 5 Dimensions 103 6 1 Unit mm A 3 01 10 1997...

Страница 6: ...CH3 5145 8 DESCRIPTION OF MECHANISM 1 Light Path Diagram path From objective...

Страница 7: ...The interpupillary mo ing the left and right prisms in parallel distance is adjusted by moving these to each other prisms in an arc centering the optical axis Less possibility of optical axis displace...

Страница 8: ...CH3 Bl45 B DESCRIPTION OF MECHANISM 3 lnterpupillary Distance Adjustment Mechanism 3 l Working force The interpupillary distance working force is adjusted by screwing this part B 3 01 10 1997...

Страница 9: ...f the sleeve Ali the optica axis with the mechanicalrotation center by changing the sleeveposition so that the image in the field of view doesnot move upon observation The left sleeve is used to adjus...

Страница 10: ...Bl45 B DESCRIPTION OF MECHANISM 4 2 Tube length Left sleeve Right sleeve Adjust 0 position of helicoid scaleto the The tube length can be changedby adjusting this index sleeve Rvertically f B 5 01 10...

Страница 11: ...ical axis imagesurfacewhen the combinethe standardeyepiece KNOO48 the right sleeve interpupillary distanceis microscopeframe product and the standard only 62 mm objective KNOO15 Then check the displac...

Страница 12: ...sleeve periphery and pull with the tension gauge Checkino the left right optical axis I KN0048 Objective 4x or 10x Cross micrometer or Test plate Note Whenusingthetensiongaugeto measure in theopensta...

Страница 13: ...CH3 Bl45 C INSPECTION STANDARD FT 36 Checking the tube length Checking the exit pupil center X Use the IWO048 accessory adapter tupsidedown Objective 4x or 10x c 3 01 10 1997...

Страница 14: ...CH3 B145 D DISASSEMBLY AND ASSEMBLY PROCEDURES 1 Entire Disassembly and Assembly 2 b a m 6 D l 01 10 1997...

Страница 15: ...ACU3X3SA 5 2PCS Refer to E 15 for the repair procedure SLEEVE R ASS Y CUK3X6SB 6 2 PCS Refer to E 5 for the repair procedure SLEEVE R ACU3X3SA 7 2 PCS Refer to E 14 for the repair procedure t OC MOUK...

Страница 16: ...Adhesive Remarks iI C RING SHAFT FIXNGPLATE PRISM DIAPHRAGM PRISM OT1595 CUK3X6SA 11 4 PCS Refer to E 10 for the repair procedure CUTS3X6SA 12 1 pc OT1873 To be pushed all the way in the arrow directi...

Страница 17: ...0 in the following procedure 1 lMiGmize the interpupillary distance 2 Stick the DIVISION PIATE so that the ROUND PROJECTION of the sleeve R assembly and the ROUND PROJECTION of the helicoid ring assem...

Страница 18: ...length Rightsleeve movethesleeve Rvertically 1 Left sleeve changethehelicoidscaleposition I ROCEDURE f 1 Adjust the revolving axis by changing the right sleeve position f Refer to E 2 2 Adjust the le...

Страница 19: ...lary f distance by fully opening the i right sleeve see the illustration Objective 4x or 10x Test plate 3100 mm concentriccircles Align the specimen center with the cross hairs center of the standard...

Страница 20: ...s reduced the concentric circles center moves from B to A The revolving axis center 0 is located the left hand side of the moving direction When the interpupillary distance is increas ed the revolving...

Страница 21: ...imen centerwith the revolving axis center found in 2 above Work Clove the spec mencenter from A to 0 by turning hestage control knob m 3 Kc JFizie JJ II 4 a 3 w I Image seen through the standard eyepi...

Страница 22: ...wo screws slightly which 1Image at first secure the right sleeve Use a thin screwdriver A thick screwdriver i i may damagethe outside surface of f I i 2 Align the cross hairs center of the 1The center...

Страница 23: ...eck if the specimencenter in the right sleeveis seenwithin therangeshown in Fig B Fig A i y __ the right sleeveshouldbe see 2 divisions Magnifiedview of the center Thecenter positionis within the ihad...

Страница 24: ...ttheinterpupillarydistanceto about62 mm Seethe illustration _ Test plate j 100 mmconcentric circles Positionof about62 mm Align the specimen center with the cross hairs center of the standard eyepiece...

Страница 25: ...enterof thestandardeyepieceandthe crosscenterof the pecimenaredisplaced Align the cross hairs center of the standard eyepiece with the cross center of the specimen Work Loosen the nvo screwssli tly w...

Страница 26: ...Fig A and check if the specimen center in the left sleeve is seenwithin the range shown in Fig B Fig A i Fig B i The specimencenter in the left sleeve should be seer within theshadedarea Magnified vi...

Страница 27: ...crewsslightly whichsecure the BI unit to the extentthattheunit does not becomeloose _ J Setthe interpupillary distanceto about62 mm Position of about62 mm 7 y iI iC iS i 1 Image seen through the stand...

Страница 28: ...ersarealigned _ le carefulnot to disturbthe adjustment HECKNG THE ABSOLUTE f Change the interpupillary distance as shown in Fig A and check PTICAL AXIS i the specimen center in the right sleeve is see...

Страница 29: ...xit pupil diameter and the cross Check that the displacementbetweenthe centerof objective s exit pupil diameter and the cross hairs center of centering telescopeis within 20 of the objective s exit pu...

Страница 30: ...CH3 Bl45 E REPAIR PROCEDURE f 1 5 Adjustingthe tube length X Insertthe standardeyepieceinto the right sleeve KNOO3S are clearlyseen E 13...

Страница 31: ...throughthe standardeyepiece 3 Tighten the screwswhichwereloosenedin 1 X Don t tighten thescrewstoo strongly The sleeve maybe deformed Turn thescrewabout60 cloch e after thescrewendcontactsthesleeve 6...

Страница 32: ...by turningthehelicoidring of left sleeveso that The fine stripesin thebackgroundof the 3 Loosenthetwo screwswhichsecuretheSCALE RING Whenlooseningthescrew be carefulnot i to disturbthe adjustment ACU...

Страница 33: ...different they must be within the standardrang2 shown in Fig A Fig A Standardrange indicatedby scale i The clearly seenpositiondifferencebetweenthe i cross hairs of the standard eyepiece and the i spe...

Страница 34: ...the ADJ SCREW 0 in the direction A to increase the working force Turn the ADJ SCREW 0 in the direction B to decreasethe working force 2 3 Tighten the screwsof theADJ SCREW0 to fix the working force f...

Страница 35: ...escope C l 3 E 12 KNOO48 Universal standardeyepiecewith crosshairs C l 3 E 2 7 10 13 ___ ___________ _______________________ KNO048 accessoryadapter 2 C l 3 12 OT1068 Tension gauge 3 kg Y C l 2 E 17 F...

Страница 36: ...CH3 Bl45 G LUBRICANTS AND CHEMICALS 1 List of Lubricants No OT1 595 Silicone gease Description D 2 3 Page 2 List of Chemicals No Description OTB73 Silicone adhesive black D 3 Page G l...