DMTA048-01EN, Rev. C, February 2013
52 Chapter 4
CFU-PWZ: Remote
control’s light is not
on when the remote
control’s power
button is pressed
down and the
pump’s motor does
not turn on.
Bad electric
connection
Check these connections:
Power plug in AC power outlet
(wall)
•
power cable in the power
supply unit’s AC In
connection
•
power supply connector
in the remote control’s
DC
IN
receptacle
Table 4 Troubleshooting
(continued)
Problem Description
Possible Cause
Solution