Important Information — Please Read Before Use
EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL
• Never insert or withdraw the endoscope’s insertion tube
while the bending section is locked in position. Patient injury,
bleeding and/or perforation can result.
• Never perform flexibility adjustment, operate the bending
section, feed air or perform suction, insert or withdraw the
endoscope’s insertion tube, or use endo-therapy accessories
without viewing the endoscopic image. Patient injury,
bleeding and/or perforation can result.
• Never perform flexibility adjustment, operate the bending
section, feed air or perform suction, insert or withdraw the
endoscope’s insertion tube, or use endo-therapy accessories
while the image is frozen. Patient injury, bleeding and/or
perforation can result.
• Regardless of the flexibility of the endoscope’s insertion tube,
never insert or withdraw the insertion tube abruptly or with
excessive force. Patient injury, bleeding and/or perforation
can result.
• Never insert or withdraw the endoscope’s insertion tube, use
endo-therapy accessories while the image is magnified.
Patient injury, bleeding and/or perforation can result (only
when using the image magnification of the video system
center CV-180).
• Do not touch the light guide of the endoscope connector
immediately after removing it from the light source because it
is extremely hot. Operator or patient burns can result.
• When the endoscopic image does not appear on the monitor,
the CCD may have been damaged. Turn the video system
center OFF immediately. Continued power supply in such a
case will cause the distal end to become hot and could cause
operator and/or patient burns.
• If it is difficult to insert the endoscope, do not forcibly insert
the endoscope; stop the endoscopy. Forcible insertion can
result in patient injury, bleeding and/or perforation.
• When combining the endoscope with a splinting tube, there is
the risk of perforation or bleeding due to entanglement of the
mucous membrane, or of the tube to becoming separated
from the endoscope and remaining in the body. Before use,
be sure to read the instruction manual for the splinting tube to
fully understand its characteristics.