2
BXFM
Illuminator
Mounting Holder
BXFM-ILHS
BXFM-ILH
Standard
Combination
Weight (Module
Names)
7.5 kg
BX-RLA2, U-TR30-2,
U-LH100-3, revolving
nosepiece, objec-
tives, eyepiece
Additional Load
Limit
2.8 kg
4.0 kg
When the BXFM-
ILHSPU counter
spring is used
5.2 kg
U-KMAS, U-TR30-2,
U-LH100-3, revolving
nosepiece, objec-
tives, eyepiece
· With the combination using the BXFM-ILH illuminator mounting holder,
do not use the U-ST stand because this makes the system unstable.
· The system can be stabilized by using the SZ-STL large stand, but the
image may fluctuate or blur irregularly when a 50X or higher-power ob-
jective is used.
· It is recommended to fabricate an exclusive clamping tool which uses
the support mounting hole (P. 20) or module mounting screw hole of the
BXFM-F focusing unit.
Symbol
Explanation
The following symbols are found on the microscope. Study the meaning of the symbols and always use the equip-
ment in the safest possible manner.
Indicates that the surface becomes hot, and should not be touched with bare hands.
Before use, carefully read the instruction manual. Improper use could result in personal injury to
the user and/or damage to the equipment.
Indicates that the main switch is ON.
Indicates that the main switch is OFF.
Warning engraving/stickers are placed at parts where special precaution is required when handling and using the
microscope. Always heed the warnings.
Warning engraving
position
Lamp housing
(Warning against high temperature)
l
Safety Symbols
Warnings
9. When using a photographic unit or TV camera, be sure to observe the
total weight limit of the modules added to the standard combination as
shown below.
Содержание BXFM Series
Страница 2: ......
Страница 24: ...21 Provided with Illuminator Mounting Holder Counter Spring BXFM F BXFM ILH BXFM ILHSPU ...
Страница 33: ...MEMO ...
Страница 34: ...MEMO ...