BX2 MAINTENANCE MANUAL
MAINTENANCE PROCEDURE
- 2 -
2. Guide to maintenance
2-1 Overview of maintenance
Set your correct interpupillary distance.
Note any areas suggesting a need for
mechanical and/or optical maintenance
by operating it or observing a specimen.
Sweep off dust on the outer surfaces
with the soft brush. If there are stains
on the microscope frame, clean them
with neutral detergent.
Remove the revolving nosepiece, stage
and condenser.
Wipe off any dirt spots on the surface.
The plastic parts should be cleaned with
neutral detergent.
Clean all exposed optics such as
lenses of eyepieces and objectives.
(Only the BX41collector lens should
be cleaned with neutral detergent
because it is made of plastic.)
Mount the removed components to
the microscope frame.
Polish all plastic components and the
frame with silicon cloth.
Return the interpupillary distance to
original condition and prepare for
the inspection. (see page 6)
Do a final check following the inspection
sheet in this manual. (See page 7)
Tool required:
1)
2)
3)
7)
4)
5)
6)
8)
9)
Lens tissue
Neutral detergent
(For plastic part or frame)
Cotton swab
Blower
Silicon cloth
(For finishing)
Cleaning solution
(For lens or filter etc.)
Plastic part
Plastic part
collector lens
or tweezers etc.
OL YMPUS
BX51
PRE-SET
12
11
10
9
8
6
4
P
OL YMPUS
BX51
PRE-SET
12
11
10
9
8
6
4
P
OL YMPUS
BX51
PRE-SET
12
11
10
9
8
6
4
P
OL YMPUS
BX51
PRE-SET
12
11
10
9
8
6
4
P
OL YMPUS
BX51
PRE-SET
12
11
10
9
8
6
4
P
OLYMPUS
BX51
PRE-SET
12
11
10
9
8
6
4
P