![Olympus BX-RLA2 Скачать руководство пользователя страница 25](http://html1.mh-extra.com/html/olympus/bx-rla2/bx-rla2_instructions-manual_3978253025.webp)
BX-RLA2
22
Optical
characteristics
Magnifi-
cation
N.A.
W.D.
(mm)
Cover
glass
thick
ness
(mm)
Resolu-
tion
(µm)
Eyepieces
Series
WHN10X (FN22)
SWH10X (FN26.5)
Total
mag.
Depth
of focus
(µm)
Field
of view
(mm)
Total
mag.
Depth
of focus
(µm)
Field
of view
(mm)
MPLFLN
Plan Semi-
Apochromat
(FN26.5)
*
1.25X:FN22
MPLFLN-BD
Brightfield/
darkfield
Plan Semi-
Apochromat
(FN26.5)
1.25X 0.04
3.5
—
8.39
12.5X
870
17.6
—
—
—
2.5X 0.08
10.7
—
4.19
25X 220
8.8
25X
220
10.6
5X 0.15
20.0
—
2.24
50X
59
4.4
50X
59
5.3
10X 0.30
11.0
—
1.12
100X
15
2.2
100X
15
2.65
20X 0.45
3.1
0
0.75
200X
5.2
1.1
200X
5.2
1.33
50X 0.80
1.0
0
0.42
500X
1.3
0.44
500X
1.3
0.53
100X 0.90
1.0
0
0.37
1000X 0.73
0.22
1000X
0.73
0.27
5X 0.15
12.0
—
2.24
50X
59
4.4
50X
59
5.3
10X 0.30
6.5
—
1.12
100X
15
2.2
100X
15
2.65
20X 0.45
3.0
0
0.75
200X
5.2
1.1
200X
5.2
1.33
50X 0.80
1.0
0
0.42
500X
1.3
0.44
500X
1.3
0.53
100X 0.90
1.0
0
0.37
1000X 0.73
0.22
1000X
0.73
0.27
150X 0.90
1.0
0
0.37
1500X
0.6
0.15
1500X
0.6
0.18
Marking
MPlanFLN
MPlanFLN-BD
MPLFLN-BDP
Reflected Polarized
Light Plan Semi-
Apochromat
(FN26.5)
MPlanFLN-BDP
LMPLFLN
Long-WD Plan
Semi-Apochromat
(FN26.5)
LMPlanFLN
LMPLFLN-BD
Brightfield/darkfield
long-WD Plan Semi-
Apochromat
(FN26.5)
LMPlanFLN-BD
5X 0.15
12.0
—
2.24
50X
59
4.4
50X
59
5.3
10X 0.25
6.5
—
1.34
100X
18
2.2
100X
18
2.65
20X 0.40
3.0
0
0.84
200X
6.1
1.1
200X
6.1
1.33
50X 0.75
1.0
0
0.45
500X
1.4
0.44
500X
1.4
0.53
100X 0.90
1.0
0
0.37
1000X
0.73
0.22
1000X
0.73
0.27
5X 0.13
22.5
—
2.58
50X
70
4.4
50X
70
5.3
10X 0.25
21.0
—
1.34
100X
18
2.2
100X
18
2.65
20X 0.40
12.0
0
0.84
200X
6.1
1.1
200X
6.1
1.33
50X 0.50
10.6
0
0.67
500X
2.5
0.44
500X
2.5
0.53
100X 0.80
3.4
0
0.42
1000X
0.87
0.22
1000X
0.87
0.27
5X 0.13
15.0
—
2.58
50X
70
4.4
50X
70
5.3
10X 0.25
10.0
—
1.34
100X
18
2.2
100X
18
2.65
20X 0.40
12.0
0
0.84
200X
6.1
1.1
200X
6.1
1.33
50X 0.50
10.6
0
0.67
500X
2.5
0.44
500X
2.5
0.53
100X 0.80
3.3
0
0.42
1000X
0.87
0.22
1000X
0.87
0.27
Note) When an MPLN-BD series objective is used in darkfield observation with a xenon light source, the peripheral area
may be obscured with certain specimens.
UIS2
series