Olympus BL-AC10DS-A2 Скачать руководство пользователя страница 3

 

 
 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

4

     

Attaching the cantilever chips to your instrument

       

3) In opening the case, put the plastic case label-side down on a desk. 

The cantilevers are tip-side-up as viewed in the case. 

4) Open the case. 

 

1)

 

Please read the instruction manual of your scanning probe microscope before this operation. 

 
 

This cantilever can not be used at a commercially SPM due to its smallness, 9 

m(L) x 2 

m(W) x 

0.13 

m(T). 

 
 
 
 
 

To attach the cantilever chip to the chip holder in your SPM, 

2) Place the chip on the adhesive stick gently against the prescribed part in your chip holder. 

(see left below) 

3) Press the center of the chip with sharp pointed tweezers and pull apart the stick from the chip. 

(see right below) 

3

     

Picking up the cantilever chip from the case

         

1) Press down gently on the part between the center and one third of the chip along the 

center line with the adhesive stick (see left below). 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

Avoid any contact with the cantilevers when you pull up the cantilever 
tip from the case. 

Caution

4) Lock the chip into the chip holder in accordance with the manner described in the manual of 

your SPM. Most of the chip holders in AFMs use a leaf spring or a wire for pressing the chip 
to the holder. It is recommended that the chip is pressed with those at the center of the chip 
to achieve a good mechanical coupling between the chip and the holder. 

 

Cantilever chips should always be placed tip-side-up. 

Caution

2) Do not press the chip too much, 

or the chip might be buried in 
the Gel sheet and the cantilever 
must break. 

 
3) Pull up the stick carefully, then 

the adhered chip is picked up 
together (see below). 

Note : Some of both-side adhesive tapes have not enough adhesive to pick up the chip 

from the Gel sheet. Then use the tweezers for this procedure. 

Note) To avoid contact with the cantilever, 

the tweezers should not access to the 
chip from cantilever side. Put the 
tweezers on the chip like following 
illustration. 

- 6e - 

- 5e - 

Содержание BL-AC10DS-A2

Страница 1: ...contained in the plastic case Tac carrier Cantilever One cantilever elongates from the side of round shoulders of the chip Magnified illust Lot Number Inspection stamp Caution Please handle our cantilevers carefully because they are fragile Caution Do not drop or shake the cantilever case Even when the cantilever chips are contained in the cantilever case the cantilevers may break if the case is h...

Страница 2: ...ike Toothpick Both side adhesive tape Anti static mat Use of tweezers made of insulator is recommended Use of anti electrostatic discharge mat and a wrist band is recommended Stereo microscope Please handle our cantilevers carefully because they are fragile Caution 4e 7e 3 Wind both side adhesive tape around the top of the thin stick to prepare an adhesive stick for picking up the chips from the c...

Страница 3: ... stick see left below Avoid any contact with the cantilevers when you pull up the cantilever tip from the case Caution 4 Lock the chip into the chip holder in accordance with the manner described in the manual of your SPM Most of the chip holders in AFMs use a leaf spring or a wire for pressing the chip to the holder It is recommended that the chip is pressed with those at the center of the chip t...

Страница 4: ...mmended to fabricate another thin wire tip like a carbon at the apex of the tip Ver 1 1 Jul 7 2010 OLYMPUS CORPORATION Please contact following if you have any question on this user s manual OLYMPUS CORPORATION Microtechnology R D Division 2 3 Kuboyama cho Hachioji shi Tokyo 192 8512 Japan Email probe olympus co jp Please access to the web page of OLYMPUS micro cantilevers http www olympus co jp p...

Отзывы: