Complete Teardown, Cleaning, and Reassembly of the Olympus BH2-5RE Modular Revolving Nosepiece                

   

 

Page 15 of 18 

 

 

Revision 3 

Reinstall the Knurled-Rubber Grip Ring  

The  final  assembly  step  for  the  turret  assembly  is  to 
reinstall  the  black  knurled-rubber  grip  ring  around  the 
outer perimeter of the revolving turret and secure it in 
place with a suitable color-matching adhesive. 
Black  E6000  is  a  good  choice  of  adhesive  here,  since  it 
will bond well with the metal of the revolving turret as 
well  as  to  the  black  knurled-rubber  grip  ring.  
Additionally,  any  of  the  inevitable  squeeze-out  that  is 
not completely removed will not be visible.  Apply two 
dabs of the black adhesive (item S1 of 

Appendix 1

onto 

the  outer  perimeter  of  the  revolving  turret,  spaced 
approximately 180° apart (see 

Figure 42

). 

 

Figure 42 

 Apply black adhesive to revolving turret 

Carefully reinstall the knurled-rubber grip ring onto the 
outer perimeter of the revolving turret, making sure to 
not stretch or damage the grip ring in the process (see 

Figure 43

). 

 

Figure 43 

 Reinstall the knurled-rubber grip ring 

Use  dry  cotton  swabs  lightly  dampened  with  isopropyl 
alcohol  to  thoroughly  remove  any  adhesive    squeeze-
out (see 

Figure 44

). 

 

Figure 44 

 Remove any adhesive squeeze-out 

Ready for Service 

The  newly  reconditioned  BH2-5RE  modular  revolving 
nosepiece assembly is now ready to provide many more 
years of service (se

Figure 45

)

 

Figure 45 

 BH2-5RE ready to be put back into service 

Requirements for Periodic Maintenance 

Periodic cleaning and application of fresh grease to the 
mechanical  detent  notches  in  the  stationary  base  is 
necessary to minimize wear of the mechanical detents, 
thereby  maximizing  the  useful  service  life  of  the 
BH2-5RE.    This  can  be  easily  accomplished  by  simply 
removing the protective cover, cleaning and re-greasing 
the detent notches, and then reinstalling the protective 
cover  per  the  procedures  detailed  in  this  document.  
The  dovetail  slide  does  not  need  to  be  removed  to 
perform  this  periodic  maintenance.    If  the  equipment 
sees heavy usage, this service should be performed on a 
six-month interval

3

.   

 

3

  Field  experience  has  shown  that  even  heavily  used  scopes  (such  as  those 

used  in  hospitals  and  clinical  lab  settings),  when  lubricated  on  a  six-month 
interval, can be expected to provide many years of trouble-free service. 

Содержание BH2-5RE

Страница 1: ...ribute this PDF document in its entirety for personal non commercial purposes only provided that the contents are not modified in any way including the copyright notices contained herein This document...

Страница 2: ...onary Base from the Turret 10 Clean Grease from the Various Components 11 Reinstall Stationary Base into Revolving Turret 11 Apply Ring of Grease for the Perimeter Balls 11 Reinstall the Perimeter Bea...

Страница 3: ...0 Figure 20 Remove the threaded retaining ring 10 Figure 21 Remove the perimeter bearing balls 10 Figure 22 Lift and remove the stationary base 10 Figure 23 Reinstall stationary base into revolving tu...

Страница 4: ...lympus BH2 5RE Modular Revolving Nosepiece Page 4 of 18 Revision 3 Figure 45 BH2 5RE ready to be put back into service 15 Figure 46 Revolving turret with severely worn mechanical detents 17 Figure 47...

Страница 5: ...on revolving modular nosepiece assembly used on the Olympus BHS BHSP BHT and BHTP microscope stands This procedure is also applicable the BH2 6RE six position modular revolving nosepiece assembly due...

Страница 6: ...due to solvent absorption The list of solvents generally considered safe for nylon includes acetone diethyl ether heptane mineral spirits naphthalene and turpentine Before Starting the Overhaul Remove...

Страница 7: ...ve unless the proper tool is used Do not attempt to remove the slotted lock ring from the pivot adjustment screw unless you have access to such a tool since the slot in the lock ring will likely be da...

Страница 8: ...causing any damage to the fragile brass threads in the revolving turret Figure 10 Insert screwdriver into bores to lock turret Loosen and Remove the Slotted Lock Ring Hold the turret assembly by gras...

Страница 9: ...taked in place with adhesive and can be surprisingly stubborn to remove so be sure to use the proper JIS screwdriver to prevent damaging the screw heads It might also be helpful to heat the screws wit...

Страница 10: ...ter punch or nailset tool into one of the five detent notches in the revolving turret instead of into one of the four tooling holes in the threaded retaining ring or the revolving turret will be irrep...

Страница 11: ...of the revolving turret see Figure 24 Do not apply too much grease here to minimize the squeeze out that will occur when the threaded retaining ring is reinstalled to hold the stationary base into the...

Страница 12: ...that the center pivot bore has been greased reinstall the bearing ball into the freshly greased center pivot bore The grease will hold the center pivot bearing ball in the proper position during subs...

Страница 13: ...with the other hand The motion of the turret should not feel gritty erratic or excessively stiff If it does loosen the slotted lock ring readjust the pivot adjustment screw and retighten the slotted...

Страница 14: ...suitable JIS screwdriver to reinstall three M2X3 countersink screws to secure the protective cover onto the stationary base see Figure 39 Figure 39 Secure protective cover using three screws Reinstall...

Страница 15: ...rubber grip ring Use dry cotton swabs lightly dampened with isopropyl alcohol to thoroughly remove any adhesive squeeze out see Figure 44 Figure 44 Remove any adhesive squeeze out Ready for Service Th...

Страница 16: ...f the stationary base If there is excess play in the center pivot ball such that the turret is not held in the exact center point the whole turret may move slightly resulting in poor objective centeri...

Страница 17: ...lored plating on the exposed surface When the hard plating in the area of the mechanical detents begin to wear away as a result of severe use without routine cleaning and re greasing of the detents th...

Страница 18: ...document Item Description Manufacturer Manufacturer Model Vendor Vendor Price S1 E6000 adhesive black 2 oz E6000 Amazon 8 22 S2 Plastilube Brake Grease 75 cc Plastilube ATE70015 Amazon 10 88 Plastilu...