background image

A

B

C

D

1

2

3

4

D

C

B

A

4

3

2

1

REV

REVISION DESCRIPTION

DATE

ECO #

A

A

A

A

A

A

A

A

A

A

R212

2k

1/8W

Q202

2SA817

+ C201

47uF
25V

R209

2k

1/8W

D203

1N4148

D201

1N4002

+

R204

820

1/8W

R201

0.1 OHM

5W

3

2

11

1

4

+

-

IC1-A

uPC324C
NEC

R203

820

1/8W

LED

"B"

2V

R211

1.8k

1/8W

12

13

14

IC1-D

uPC324C
NEC

R207

820

1/8W

5

6

7

IC1-B

uPC324C
NEC

LED

"E"

7V

R210

1.5k

1/8W

D202

1N4148

D204

1N4148

LED

"A"

POWER

10

9

8

IC1-C

uPC324C
NEC

R213

10k

1/8W

R214

10k

1/8W

RV101

500

OHM

NOBLE

Q101

2SD867

J101

DB101

LED

"C"

4V

R205

820

1/8W

LED

"D"

6V

R202

200 Ohm

1/8W

R208

4.3k

1/8W

R206

820

1/8W

Q201

2SA817

T101

F101

+

C202

.047uF

R5

500

OHM

BLUE

BLACK

BLUE

BROWN

BLACK

BLACK

BLUE

BLACK

BLACK

(ONE OF FOUR)

AS PUBLISHED IN THE

OLYMPUS REPAIR MANUAL

RED

POWER

TRANSISTOR

YELLOW

VIOLET

ORANGE

RED

BLUE

BLUE

RED

VIOLET

BLACK

YELLOW

100V (220V)

BLACK

BLACK

J103

VIOLET

115V (240V)

WHITE

BLACK

WHITE

VIOLET

BLACK

BLUE

GROUND

BROWN

NEUTRAL

SEE NOTE 1

LINE

CM-3

GREEN / YELLOW

AC INLET

B

C201

BLACK

WHITE

BLACK

WHITE

POWER TRANSFORMER

BLACK

- APPENDIX 3 - 

OLYMPUS BH-2 (LATER)

BHT/BHTU ELECTRONICS

4- THE AVERAGE DC VOLTAGE ON THIS NODE VARIES FROM APPROXI10.5V (AT MINIMUM LAMP VOLTAGE)
     TO +7.6V (AT MAXIMUM LAMP VOLTAGE), DUE TO INCREASED RIPPLE AT HIGHER CURRENT DRAINS.

RED

- DETAIL 3 -

PINOUT OF BRIDGE

RECTIFIER DB101

-INSTALL WITH THERMAL PASTE-

(VIEW FROM TOP OF BASE)

D202

SEE NOTE 5

BLACK

2- THE UYPC48 PCB SCHEMATIC ON PAGE 98 OF THE OLYMPUS BH2(BHS) REPAIR MANUAL DIFFERS FROM THE SCHEMATIC
     PRESENTED HERE IN THE WIRING OF THE BAR-GRAPH LEDS AND THEIR CURRENT-LIMITING RESISTORS.  SEE DETAIL 1.

3- THE UYPC48 PCB SCHEMATIC ON PAGE 98 OF THE OLYMPUS BH2(BHS) REPAIR MANUAL DIFFERS FROM THE SCHEMATIC
     PRESENTED HERE IN THE WIRING OF THE CONTROL VOLTAGE CLAMP.  SEE DETAIL 2.

NOTES:

AC

IN

BROWN

1

3-10-16

HALOGEN LAMP

6V/20W, G4 BASE
OLYMPUS 8-C405

OLYMPUS UYPC48

1- FUSE F101 IS PRESENT ONLY ON THE 240V VERSION OF THE BHTU STAND.

SEE

NOTE 6

6

8

AC

IN

7

J1

3

D204

D201

4

5

2

- DETAIL 1 -

DISCREPANCY WITH OLYMPUS SCHEMATIC

- BAR-GRAPH LEDS -

SEE NOTE 2

1

D202

S15VB10

REAR OF BASE

SEE DETAIL 3

1

C201

FILTERED

DC (+)

SUPPLY

P/O P1

DB101 (+)

Initial Release

DT 0103 (115V)
DT 0104 (240V)

FILTERED

DC SUPPLY

(+)

BLUE

DB101 (+)

(ONE OF FOUR)

ACTUAL CIRCUITRY

ON THE UYPC48 BOARD

ACTUAL CIRCUITRY

ON THE UYPC48 BOARD

SEE DETAIL 4

E

SEE

NOTE 2

C

YELLOW

+7.6 TO +10.5 VDC  SEET NOTE 4

SEE

NOTE 3

ALPS SDT-7

POWER SWITCH

SW101

REAR OF BASE

---

VOLTAGE

SELECT

SW102

D201

D203

R214

FILTERED

DC SUPPLY

(+)

BLACK

R214

P/O P1

- DETAIL 4 -

PINOUT OF POWER

TRANSISTOR Q101

-INSTALL WITH THERMAL PASTE-

(VIEW FROM TOP OF BASE)

NEG
OUT

D204

FILTERED

DC (+)

SUPPLY

AS PUBLISHED

IN THE OLYMPUS

REPAIR MANUAL

BLACK

BLUE

AND

BROWN

POS
OUT

D203

RED
AND

BLACK

2

BRT

COPYRIGHT (C) 2016 CARL HUNSINGER

ORANGE

ORANGE

4

BRT

J2

J1

DV2288-01 LIGHT PRESET BOARD

RED

INTENSITY

SLIDER

2

REVISION 1

5- SEE APPENDIX 2 FOR EARLY PRODUCTION UNITS THAT DO NOT HAVE A LIGHT PRESET CONTROL / SWITCH.

- DETAIL 2 -

DISCREPANCY WITH OLYMPUS SCHEMATIC

- CONTROL-SIGNAL CLAMP -

SEE NOTE 3

RED

ORANGE

LATER PRODUCTION UNITS WITH
LIGHT PRESET CONTROL / SWITCH

7- W201, W202, W203, AND W204 (NOT SHOWN IN THE SCHEMATIC) ARE ZERO-OHM CIRCUIT JUMPERS INCLUDED TO ALLOW
     A SINGLE LAYER PCB. THESE COMPONENTS ARE PRESENT ON EARLY PCBS ONLY.

3

6- NEVER OPERATE THE MICROSCOPE WITHOUT THE CENTER TERMINAL OF THE AC INLET ELECTRICALLY BONDED TO THE CHASSIS.

LAMP HOUSE

LS20H (OLYMPUS 5-LB402) OR

LS20H-M2 (OLYMPUS 5-S119)

1

Содержание BH-2 BHT

Страница 1: ...14 2016 Carl Hunsinger All rights reserved This document may not be reproduced or distributed in whole or in part for any commercial purpose without the express written consent of the author Permissio...

Страница 2: ...t jack on the back where the line cord connects There is a power switch on the front to turn the illumination on and off as well as an intensity control on the right hand side to vary the lamp voltage...

Страница 3: ...ward biasing its base emitter junction and allowing it to increasingly conduct collector current Since the collector current of transistor Q201 is the base current of transistor Q101 Q101 in turn begi...

Страница 4: ...setting of the intensity control and with an abnormally low lamp resistance of 0 2 LED Bar Graph Display Circuitry Figure 3 is a simplified schematic diagram of the LED bar graph display circuitry Thi...

Страница 5: ...comparators were allowed to simultaneously drop to zero the comparators would be in an indeterminate state and depending on the input offset voltage parameters of the specific op amp devices their out...

Страница 6: ...Olympus repair manual would not guarantee that the LEDs remain off around the zero crossings of the AC line Removing the Electrical Base from the Stand To gain access to the electronics in the base o...

Страница 7: ...may be damaged if the stand is not placed on a suitable clutter free surface Store the stand someplace out of harm s way and where dust and debris will not accumulate on the exposed lighting component...

Страница 8: ...exposed points within the electrical base Plug the AC line cord into a properly grounded AC receptacle The electronics may now be operated as necessary for troubleshooting purposes Never operate the m...

Страница 9: ...r into the control so as to minimize overspray and blow back onto nearby components Be sure to allow the contact cleaner sufficient time to evaporate out of the control before re applying power to the...

Страница 10: ...chassis to facilitate heat flow The NTE5322 is available from Allied Electronics and from Amazon Figure 8 Bridge Rectifier Mounted to BHT BHTU Electrical Base 100 115V Version Lamp Intensity Control T...

Страница 11: ...trical base into an earlier stand that does not have the chassis notches to accommodate this board Removing and Bypassing the Light Preset Board Remove the light preset board from the electrical base...

Страница 12: ...PDT slide switch rated for 4A 125VAC and 2A 250VAC The slide protrudes 3 8 from the mounting surface and the mounting holes are spaced 1 1 8 on center There are openings on both ends of this switch th...

Страница 13: ...normal operation and because of this the solder joints can sometimes go bad Resoldering the leads of R201 may be enough to fix an inoperative lamp dimmer Resistor R201 can be replaced with a Vishay CP...

Страница 14: ...parts listed above can be found look for an NPN part with a collector current rating of at least 15A a collector to emitter voltage rating of at least 100V and a current gain of at least 40 in a suit...

Страница 15: ...by Olympus as part number 5 LB402 did not have a metal reflector behind the bulb Later in the life of the BHT BHTU Olympus upgraded the lamp house to the LS20H M2 part number 5 S119 which incorporated...

Страница 16: ...a replacement for the back cover and a new G4 lamp socket may be used to overhaul either the LS20H or the LS20H M2 lamp houses The replacement back is provided since the original can be damaged durin...

Страница 17: ...n the new socket to match those of the original socket and solder the new socket in place Reinstall the two screws to secure the new socket in its proper position Reinstall the back cover securing wit...

Страница 18: ...wn relative to the negative output of the bridge rectifier which is the zero reference point These waveforms were produced using version 4 19e of the LTSpice IV circuit simulator Figure 21 Lamp Wavefo...

Страница 19: ...Revision 2 Page 19 of 24 Lamp Dimmer Waveforms continued Figure 23 Lamp Waveforms at Low Intensity 20 Setting of Intensity Control Figure 24 Lamp Waveforms during Current Limiting 80 Setting of Inten...

Страница 20: ...U Electronics Revision 2 Page 20 of 24 LED Bar Graph Display Waveforms Figure 25 Filtered DC Power Supply Line at Maximum Lamp Intensity 100 Setting of Intensity Control Figure 26 Bar Graph Comparator...

Страница 21: ...e 21 of 24 LED Bar Graph Display Waveforms continued Figure 27 Clamped Control Signal to Comparators at Medium Lamp Intensity 50 Setting of Intensity Control Figure 28 Comparator IC1 B Inputs with 6V...

Страница 22: ...lamp socket Item BW 990 Price 16 75 Philips 7388 halogen lamp Item BW ESB Price 6 51 Caig Laboratories Inc URL www caig com Telephone 858 436 8388 DeoxIT D5 Spray Contact Cleaner Item D5S 6 Price 18 9...

Страница 23: ...BH2 BHS REPAIR MANUAL DIFFERS FROM THE SCHEMATIC PRESENTED HERE IN THE WIRING OF THE CONTROL VOLTAGE CLAMP SEE DETAIL 2 NOTES AC IN BROWN INTENSITY SLIDER COPYRIGHT C 2014 CARL HUNSINGER HALOGEN LAMP...

Страница 24: ...IC ON PAGE 98 OF THE OLYMPUS BH2 BHS REPAIR MANUAL DIFFERS FROM THE SCHEMATIC PRESENTED HERE IN THE WIRING OF THE CONTROL VOLTAGE CLAMP SEE DETAIL 2 NOTES AC IN BROWN 1 3 10 16 HALOGEN LAMP 6V 20W G4...