DMTA-10022-01EN, Rev. C, January 2015
Chapter 5
60
5.2.5
Entering a Known Material Sound Velocity
When preparing to measure thicknesses on parts made of a different material, if you
know the sound velocity for the material, you can enter it directly without performing
a material sound velocity calibration procedure.
To enter a known material sound velocity
1.
In the measurement screen, press
[2nd F]
,
[CAL VEL] (VEL)
.
2.
In the
VELOCITY
screen (see Figure 5-9 on page 60), use the arrow keys to edit
the velocity to the known value.
Figure 5-9 Entering a known material sound velocity
3.
Press
[MEAS]
to return to the measurement screen.
5.2.6
Locked Calibrations
The 45MG includes a password-protected locking function to prevent changes to
setups, and prevent access to certain functions. A change to the calibration is one
action that can be locked. In such case, the message shown in Figure 5-10 on page 61
appears momentarily on the help text bar (see “Locking the Instrument” on page 136
for details).
Содержание 45MG
Страница 8: ...DMTA 10022 01EN Rev C January 2015 Table of Contents viii List of Tables 227 Index 229...
Страница 26: ...DMTA 10022 01EN Rev C January 2015 Introduction 16...
Страница 36: ...DMTA 10022 01EN Rev C January 2015 Chapter 1 26...
Страница 42: ...DMTA 10022 01EN Rev C January 2015 Chapter 2 32...
Страница 136: ...DMTA 10022 01EN Rev C January 2015 Chapter 6 126...
Страница 150: ...DMTA 10022 01EN Rev C January 2015 Chapter 7 140...
Страница 157: ...DMTA 10022 01EN Rev C January 2015 Configuring the Instrument 147 Figure 8 3 The COMM screen...
Страница 158: ...DMTA 10022 01EN Rev C January 2015 Chapter 8 148...
Страница 196: ...DMTA 10022 01EN Rev C January 2015 Chapter 10 186...
Страница 208: ...DMTA 10022 01EN Rev C January 2015 Chapter 11 198...
Страница 226: ...DMTA 10022 01EN Rev C January 2015 Appendix A 216...
Страница 232: ...DMTA 10022 01EN Rev C January 2015 Appendix C 222...