The 45MG can generate inspection data reports without having to be connected to a
computer or printer. The following reports are available:
File summary
Shows basic statistics for the file (minimum thickness and location, maximum
thickness and location, and high and low alarm conditions along with the mean,
median, and standard deviation).
Minimum/Maximum summary
Shows a list of ID number locations that have the minimum and maximum
thickness in a file.
Alarm summary
Shows a list of all the ID number locations where a low alarm and high alarm
occurred.
File comparison
Allows you to select two files and compare them. The first file contains the
previous inspection data, and the second file contains the current inspection data.
The report indicates the maximum wall loss and any area of greater wall
thickness (growth), and their ID number locations.
Minimum Review
Allows you to select a file, and then review all the minimum thickness locations in
the file. In addition to verifying the thickness at all minimum locations, you can
replace these minimum thickness locations if necessary.
To generate a report
1.
In the measurement screen, press
[FILE]
, and then select
REPORT
.
2.
In the submenu, select the desired type of report. If you select:
•
FILE SUMMARY
, go to step 3
•
MIN/MAX SUMMARY
, go to step 4
•
FILE COMPARISON
, go to step 5
•
ALARM SUMMARY
, go to step 6
•
MIN REVIEW
, go to step 7
3.
In the
FILE SUMMARY
screen (see Figure 6-39 on page 121):
a)
Select the file for which you want to create the report.
Содержание 45MG
Страница 8: ...DMTA 10022 01EN Rev C January 2015 Table of Contents viii List of Tables 227 Index 229...
Страница 26: ...DMTA 10022 01EN Rev C January 2015 Introduction 16...
Страница 36: ...DMTA 10022 01EN Rev C January 2015 Chapter 1 26...
Страница 42: ...DMTA 10022 01EN Rev C January 2015 Chapter 2 32...
Страница 136: ...DMTA 10022 01EN Rev C January 2015 Chapter 6 126...
Страница 150: ...DMTA 10022 01EN Rev C January 2015 Chapter 7 140...
Страница 157: ...DMTA 10022 01EN Rev C January 2015 Configuring the Instrument 147 Figure 8 3 The COMM screen...
Страница 158: ...DMTA 10022 01EN Rev C January 2015 Chapter 8 148...
Страница 196: ...DMTA 10022 01EN Rev C January 2015 Chapter 10 186...
Страница 208: ...DMTA 10022 01EN Rev C January 2015 Chapter 11 198...
Страница 226: ...DMTA 10022 01EN Rev C January 2015 Appendix A 216...
Страница 232: ...DMTA 10022 01EN Rev C January 2015 Appendix C 222...