![Oldham OLCT 20 Скачать руководство пользователя страница 24](http://html1.mh-extra.com/html/oldham/olct-20/olct-20_user-manual_3710886024.webp)
8
– Technical Specifications
20
Sensors
Gas type
Measure-
ment range
(ppm)
Explosion
-proof
sensor
Intrinsically
-safe
sensor
Temp.
range (°C)
% RH Accuracy (ppm) Average
service life
(months)
Resp. time
T50/T90 (s)
Storage
conditions
and time
Combustible
gases
Catalytic
0-100% LEL
-20 to +55 0-95
+/-1% LEL
(from 0-70% LEL)
40
6/15 (CH4)
(b)
AsH
3
Arsine
1.00
-20 to +40 20 - 90 +/- 0.05
18
30/120
(a)
Cl
2
Chlorine
10.0
-20 to +40 10 - 90 +/- 0.4
24
10/60
(a)
ClO
2
Chlorine
dioxide
3.00
-20 to +40 10 - 90 +/- 0.3
24
20/120
(a)
CO
Carbon
monoxide
100
300
1000
-20 to +50 15 - 90 +/- 3
(0-100 range)
40
15/40
(a)
CO
2
Carbon
dioxide
0-5% vol.
-20 to +55 0 - 95 +/- 3%
48
11/30
(a)
COCl
2
Phosgene
1.00
-20 to +40 15 - 90 +/- 0.05
12
60/180
(c)
ETO
Ethylene
oxide
30.0
-20 to +50 15 - 90 +/- 1.0
36
50/240
(a)
H
2
Hydrogen
2000
-20 to +50 15 - 90 +/- 5%
24
30/50
(a)
H
2
S
Hydrogen
sulfide
30.0
100
1000
-20 to +50 15 - 90 +/- 1.5
(0-30 range)
36
15/30
(a)
HCl
Hydrogen
chloride
30.0
100
-20 to +40 15-95 +/- 0.4
(0-30 range)
24
30/150
(a)
HCN
Hydrogen
cyanide
30.0
-20 to +40 15-95 +/- 0.3
(0-10 range)
18
30/120
(c)
HF
Hydrogen
fluoride
10.0
-10 to +30 20 - 80 +/- 5%
12
40/90
(c)
NH
3
Ammonia
100
1000
5000
-20 to +40 15 - 90 +/- 5
+/- 20
+/- 150 or 10%
24
25/70
20/60
60/180
(a)
NO
Nitric oxide
100
300
1000
-20 to +50 15 - 90 +/- 2
(0-100 range)
36
10/30
(a)
NO
2
Nitrogen
dioxide
30.0
-20 to +50 15-90 +/-0.8
24
30/60
(a)
O
2
Oxygen
0-30% vol.
-20 to +50 15 - 90 0.4% vol.
(from 15-22% O2)
28
6/15
(a)
O
3
Ozone
1.00
0 to +40
10 - 90 +/- 0.03 (from 0-
0.2 ppm) +/- 0.05
(from 0.2-1 ppm)
18
40/120
(c)
PH
3
Phosphine
1.00
-20 to +40 20 - 90 +/- 0.05
18
30/120
(a)
SiH
4
Silane
50.0
-20 to +40 20 - 95 +/- 1.0
18
25/120
(a)
SO
2
Sulfur
dioxide
10.0
30.0
100
-20 to +50 15 - 90 +/- 0.7
(0-10 range)
36
15/45
(a)
CH
3
Cl
Chloro-
methane
500
-20 to +55 20 - 95 +/- 15% (from 20-
70% PE)
40
25/90
(d)
CH
2
Cl
2
Dichloro-
methane
500
-20 to +55 20 - 95 +/- 15% (from 20-
70% PE)
40
25/90
(d)
Freon R12
1% vol.
-20 to +55 20 - 95 +/- 15% (from 20-
70% PE)
40
25/90
(d)
Freon R22
2000
-20 to +55 20 - 95 +/- 15% (from 20-
70% PE)
40
25/90
(d)
Freon R123
2000
-20 to +55 20 - 95 +/- 15% (from 20-
70% PE)
40
25/90
(d)
FX56
2000
-20 to +55 20 - 95 +/- 15% (from 20-
70% PE)
40
25/90
(d)
Freon R134a
2000
-20 to +55 20 - 95 +/- 15% (from 20-
70% PE)
40
25/90
(d)
Freon R11
1% vol.
-20 to +55 20 - 95 +/- 15% (from 20-
70% PE)
40
25/90
(d)
Freon R23
1% vol.
-20 to +55 20 - 95 +/- 15% (from 20-
70% PE)
40
25/90
(d)
Freon R143a
2000
-20 to +55 20 - 95 +/- 15% (from 20-
70% PE)
40
25/90
(d)
Freon R404a
2000
-20 to +55 20 - 95 +/- 15% (from 20-
70% PE)
40
25/90
(d)
Freon R507
2000
-20 to +55 20 - 95 +/- 15% (from 20-
70% PE)
40
25/90
(d)
Содержание OLCT 20
Страница 1: ...1 User manual OLCT 20 Part Number NPO20GB Revision E 0 The Fixed Gas Detection Experts...
Страница 8: ...1 Overwiew 4...
Страница 12: ...5 Maintenance 8 2 wire Intrinsic Safety version OLCT 20i Figure 4 A B C OLDHAM controller OLCT 20 Zener Barrier...
Страница 22: ...7 Accessories 18...
Страница 28: ...8 Technical Specifications 24...
Страница 29: ...9 EU Declaration of Conformity 25 EU Declaration of Conformity...
Страница 30: ...9 EU Declaration of Conformity 26...