4
MX 15
Instruction Manual
View: Internal
Figure 2: Internal view.
Ref.
Function
See page
A.
Programming circuit (Explo 340 mA or 4-20 mA).
B.
F8 Fuse (5x20, 250 V AC - 630 mA T) with 24V DC power supply.
C.
F7 Fuse (5x20, 250 V AC - 100 mA T) with 230 VAC power supply.
D.
Secondary ground connection.
E.
Sector power supply terminal (230 V AC or 110 V AC upon request) with
the following identifications: L (phase), N (neutral) and
(ground).
F.
Power terminal block 24 V DC (0, + 24V).
G.
Terminal block connection with remote acknowledgment (dry contact
NO).
H.
Terminal block detector connection.
J.
Sensitivity settings.
K.
Zero adjustment.
L.
Configuration jumper (J17) for
Fault
relay.
- NC:
Fault
relay contact closed in alarm mode.
- NO:
Fault
relay contact open when not in alarm mode.
M.
Configuration jumper (J19) for
RL1
relay.
- NC:
RL1
relay contact closed in alarm mode.
- NO:
RL1
relay contact open when not in alarm mode.
N.
Fault
relay terminal block. Contacts CT, 250 V AC
– 2A. Contact status
not in alarm mode as defined by J17 (ref. L).
P.
RL1 alarm relay terminal block. Contacts CT, 250 V AC
– 2A. Contact
status not in alarm mode as defined by J19 (ref. M).
Q.
RL2 alarm relay terminal block. Contacts CT, 250 V AC
– 2A . Contact
status not in alarm mode as defined by J18 (ref. R).
R.
Configuration jumper (J18) for
RL2
relay.
- NC:
RL2
relay contact closed in alarm mode.
- NO:
RL2
relay contact open when not in alarm mode.