The testbench can be used to regularly check the correct operation of the MX 2100.
This system automatically injects standard gases from two cylinders installed in the testbench in order to trigger the
alarms.
The MX 2100 detects the presence of the testbench and indicates the faulty channel or channels according to the test, on
the display panel.
3
3
.
.
C
C
A
A
L
L
I
I
B
B
R
R
A
A
T
T
I
I
N
N
G
G
C
C
E
E
L
L
L
L
S
S
W
W
I
I
T
T
H
H
C
C
A
A
L
L
I
I
B
B
R
R
A
A
T
T
I
I
N
N
G
G
S
S
T
T
A
A
T
T
I
I
O
O
N
N
Calibration of the MX 2100 cells can also be performed using this automatic gas injection system which is managed by
computer and linked to the control devices via the COM2100 software.
Status and check sheets are stored on the computer.
the station is to be used in accordance with the instructions on the front panel (as in the case of the testbench).
V
V
.
.
C
C
O
O
M
M
2
2
1
1
0
0
0
0
S
S
O
O
F
F
T
T
W
W
A
A
R
R
E
E
COM2100
software ensures the appliance's supervision and maintenance:
-
display, in uncoded mode, of measurements and parameters on channels,
-
diagnostic assistance in case of failure,
-
programming of appliance and measuring channels,
-
management of options,
-
calibration of channels using automated scrolling menu,
-
output of status and check sheets,
-
management, display and printout of stored events and measurements,
-
protection by password.
The link between the MX 2100 and the PC is made by an infrared port (cord assembly), as shown opposite:
(1)
A charger or sampling unit can be used to mechanically hold the IR cord on the appliance.
Insertion of
MX 2100
Follow the indications shown on the front
of the testbench
cord
Charger or pumping unit (1)
IR component