User Manual: Easy DR
ODB-100-5070ENG
40
Rev.:17-04-2020
5
Maintenance
All maintenance to the Easy DR system is described in the relevant manufac-
turer's manuals.
Advice to consult the accompanying documentation:
•
relevant manufacturer's manuals.
No corrective maintenance should be performed by the operator, other than
solving problems as mentioned in Paragraph 5.3
5.1
Cleaning
To prevent infections, clean all parts of the equipment that are touched by the
patient (such as handgrips, chin rest, front plate), before the next patient is
examined. Do this according to local and hospital regulations, using an ethanol
or glutaraldehyde solution to disinfect and as the cleaning agent.
The front plate of the detector unit, which comes into contact with patients,
should be clean for each patient (hygiene). For example, cover the front plate
with clean sheets of paper and remove a sheet after each patient.
Alternatively, patients could wear a hospital gown.
5.2
Safety switch check
A safety switch is mounted underneath the detector unit.
Check ones a month the correct functioning of this safety switch:
Press the button to move the stand height down and press up the safety
switch.
Make sure that the downward movement of the stand is blocked.
5.3
Troubleshooting
When a fault has occurred in the connection to the PACS; the system will
display an error in the upright corner.
Because the system cannot sent the images to the PACS these examinations
will not be shown in the history list of the system, the Study list