45
3.14.2.3
Time diagram area
Two diagrams of signal to time dependence are displayed in time diagram
area (item 3, Figure 32).
3.14.2.4
Signal display area (active area)
In the display area of eddy current signals is displayed the eddy current
signal, which is represented by an image in the complex plane for the
frequency selected (item 4, Figure 32). In the active area are displayed the
signals that come into the area covered by the measuring cursor.
3.14.2.5
Bottom status bar
Figure 37
"Duration of time sweeps"
- sets the number of measurements to be
displayed on time sweeps. Duration of time sweep is measured in a range
from 256 to 32768.
“Type of measurement”
– allows to select the required type of cursor
measurement from the following list:
-
Peak-Peak
(the measurement is made between two points situated at
the max. distance from each other in cursor width);
-
Vrt-M
(vertical maximum - the measurement is made between two
points situated at the max. distance from each other in cursor width in vertical
projection);
-
Cnt-Pk
(center peak - the measurements is made between the cursor
center and the point situated at the max. distance from the center in cursor
width);
-
Cnt-K
(center cursor - the measurements is made between the center
and the cross point of cursor center signal);
ENC
(encoder)
– displays the reference number by encoder when
synchronizing from ENC.
Cursor position
– displays the current vertical position of a measuring
cursor in readings.