Figure 40. Settings - Chamber & Insert
–
Chamber selection (a -b).
Repeat this sequence to find and select the sample adapter. Scroll among different sample holders using
the arrows on the left and right sides of the screen (see pointer 1 in Figure 41 a). When the sample holder you use
appears on the screen, press
OK
to select it (see pointer 1 in Figure 41 a). The selected chamber and sample holder
will be displayed on the screen (see Figure 41 b). Press
OK
to confirm (see pointer 3 in Figure 41 b).
a
b
1
2
3
Figure 41. Settings - Chamber & Insert
–
Insert selection (a -b).
Tip
►
Chambers equipped with MW-OIL insert have been factory calibrated at a Room Temperature of 23°C
±1°C for a temperature set point of 37°C. The offset of the Chamber Base and the Chamber Lid has been
optimized to guarantee a uniform temperature for all the wells. Okolab suggests operating using factory
offset values. To view factory offset value click on Offset button (see Figure 42 a).
a
b
Figure 42. Settings - Chamber & Insert
–
Offset (a
–
b).
If the desired temperature set point could not be achieved, Okolab suggests performing Chamber calibration
to find the offset values suitable for your experimental conditions (see paragraph 8.2.5.1).