1
•
Press
‘Settings’
•
Press
‘Chamber
&
Insert’
•
Select
chamber
and
insert
in
use,
then
press
‘OK’
2
NOTE:
‘Chamber’
control
mode
uses
sensors
embedded
in
the
chamber
body.
Accuracy
+/-
0.3°C
from
set-point,
if
room
T
remains
within
+/-
0.3°C.
‘Sample’
control
mode
requires
inserting
the
provided
temperature
sensor
in
a
reference
well.
Accuracy
+/-
0.1°C
regardless
of
room
temperature.
•
Press
‘Settings’
•
Press
‘Control
Mode’
•
Select
the
Control
Mode
and
press
‘Save’
Selection of Chamber & Insert
Selection of Control Mode
Setting up UNO
Setting up UNO
5
To Humidifier
CO
2
pressure setup
Set at
1 bar
(15 psi)
Set at
2 bar
(30 psi)
3
•
Check
the
‘humidity
control
enabled’
box
•
Input
room
temperature
•
Press
‘Save’
4
•
Press
‘Enabled’
•
Press
‘OK’
NOTE: it is important to enter the right room
temperature. You may check it with the help
of the fine gauge thermocouple (green wire)
Humidity control set-up
Objective Heater (optional)
Setting up UNO
1
2
3
BASE & LID
HUMIDIFIER & HEATED TUBE
OBJECTIVE HEATER (OPTIONAL)