background image

1

• 

Press

 

 

‘Settings’

• 

Press

 

 

‘Chamber

 

&

 

Insert’

• 

Select

 

chamber

 

and

 

insert

 

in

 

use,

 

then

 

press

 

‘OK’

   

   

2

 

 

NOTE:

 

‘Chamber’

 

control

 

mode

 

uses

 

sensors

 

embedded

 

in

 

the

 

chamber

 

body.

 

Accuracy

 

+/-

 

0.3°C

 

from

 

set-point,

 

if

 

room

 

T

 

remains

 

within

 

+/-

 

0.3°C.

 

‘Sample’

 

control

 

mode

 

requires

 

inserting

 

the

 

provided

 

temperature

 

sensor

 

in

 

a

 

reference

 

well.

 

Accuracy

 

 

+/-

 

0.1°C

 

regardless

 

of

 

room

 

temperature.

• 

Press

 

 

‘Settings’

 

• 

Press

 

 

‘Control

 

Mode’

• 

Select

 

the

 

Control

 

Mode

 

and

 

press

 

‘Save’

Selection of Chamber & Insert

Selection of Control Mode

Setting up UNO

Setting up UNO

5

To Humidifier

CO

2

 pressure setup

Set at 

1 bar

 (15 psi)

Set at 

2 bar

 (30 psi)

3

• 

Check

 

the

 

‘humidity

 

control

 

enabled’

 

box

• 

Input

 

room

 

temperature

• 

Press

 

‘Save’

 

 

4

• 

Press

   

‘Enabled’

• 

Press

 

‘OK’

 

 

NOTE: it is important to enter the right room 

temperature. You may check it with the help 

of the fine gauge thermocouple (green wire)

Humidity control set-up

Objective Heater (optional)

Setting up UNO

1

2

3

BASE & LID

HUMIDIFIER & HEATED TUBE

OBJECTIVE HEATER (OPTIONAL)

Heating Devices

Отзывы: