Figure 108. T1 Calibration: start multi-point calibration (a
–
b - c).
8.
When you finish the calibration, an overview appears as in Figure 109. Click on
Save
to complete the
calibration or
Cancel
to undo.
Figure 109. T1 Multi Point Calibration.
14.3
Factory Calibration
Press
Factory Reset
icon
to cancel the last user temperature calibration and restore the factory values
(see Figure 110 a). If you are sure of the reset, press
OK,
as shown in Figure 110 b.
36.8
2
b
a
c
36.8
1
2
25.00
30.00
37.00
40.00
50.00