65
Then press
Continue
to start the next step of the calibration with Span Gas 2 and/or Span Gas 3, as shown
in Figure 86 c (after connecting the span gas to use in the next step).
Figure 86. How to set Auto Gas Calibration (a
–
b - c).
At the end of the calibration procedure, CO2 and O2
offset values are shown as in Figure 87. Press
Save
to
save the offsets or
Cancel
to undo.
Figure 87. Auto Gas Calibration completed.
13.2.4.2
Manual Gas Calibration
By pressing on
Manual
in Figure 85 b, Figure 88 a appears. Press
Continue
(see Figure 88 a) to start the
Span Calibration after checking that the proper span gas cylinder is connected to LEO.
During a Manual calibration, you can see the CO2 and O2
readings on the display.
Wait until the CO2 and O2
readings became stable.
Tip
►
Okolab suggests waiting at least 10 minutes before stopping Manual Calibration, if using Okolab
Calibration Kit.
CO2 6.00 %
O2 5.00 %
CO2 19.00 %
-
CO2 6.00 %
O2 5.00 %
CO2 4.20
O2 5.40
a
b
c
CO2 offset: 0.2 @ 6.00
O2 offset: - 0.4 @ 5.00
CO2 offset: 0.44 @ 19.00
-