45
Device1 Connector
Device2 Connector
24 VDC
Power Feeder
Figure 70. Connection of the heating devices and the power feeder
10.3.2
T sensor Connection
H401-T-CONTROLLER-OEM is equipped with a T sensor and a panel mount extension thermocouple
cable, pointer 6 and 7 in Figure 64, respectively. To use the T sensor, it is necessary to use CAB 110, (pointer 6 in
Figure 64) between the panel and the control board. To do so, follow the instructions below:
1.
Cut an opening in the panel with the size indicated in Figure 71(a).
2.
Fix the CAB 110 connector to the panel with the screws provided, pointer A (M 2,5x9) and pointer B (M
2,5x11) in Figure 71(b).
a
b
A
B
Figure 71. Installation of the panel mount extension thermocouple cable. (a) Panel opening size; (b) Fixation of the Extension
thermocouple cable
3.
Connect the extension thermocouple cable (pointer 6 in Figure 64) to the control board as shown in Figure
72(a).
4.
Connect the T sensor (pointer 7 in Figure 64) to the panel as shown in Figure 72(b).