background image

  H301-OLYMPUS-IX-SUSP 

User Manual 

   

REV 02

 

 www.oko-lab.com 

  

 

Pa

ge

8

 

4.

 

Insertion of the Sample Feedback Temperature Sensor 

Insert the Sample Feedback Temperature Sensor through the dedicated opening located in the H301-OLYMPUS-IX-SUSP 

(see Figure 8, Frontal and 3D views). 

1.Frontal view

2.3D view

Temperature Sensor dedicated opening 

Temperature Sensor dedicated opening 

 

Figure 8. Insertion of the temperature sensor inside the chamber. 

5.

 

Insertion of Sample Holder into Chamber 

Sample holders fit into the chamber base and are held in place by magnets embedded within both chamber and hold-

er.  To  introduce  a  sample  holder  with  the  proper  orientation,  match  the  red  dot  on  the  holder  to  the  one  on  the 

chamber base, as illustrated in Figure 9. 

Insert the adapter by 

superimposing the red points

 

Figure 9. Introduction of the Specimen Holder inside the Chamber Base. 

 

 

Содержание H301-OLYMPUS-IX-SUSP

Страница 1: ... www oko lab com Page 1 H301 OLYMPUS IX SUSP Compatible with the following XY stages Compatible with the following Okolab Controllers Olympus IX SUSP GX SVR GX SFR BX3 SSU IX2 SFR IX2 SVL 2 H301 T BL PLUS UNO COMBINED CONTROLLER H401 T DUAL ...

Страница 2: ...n Lid 6 3 4 Laser Interlock Lid 7 4 Insertion of the Sample Feedback Temperature Sensor 8 5 Insertion of Sample Holder into Chamber 8 6 Working with 35 or 60 mm Petri Dish Spacing Rings and Magnetic Locks 9 7 Working with 1x3 and 1x2 chamber slides magnetic locks 10 8 Connecting the Gas Supply and Filling the Humidity Module 12 9 Working with Perfusion 12 10 Connection of the Chamber with XY stage...

Страница 3: ...e of the glass lid Chamber top view Sliding Lid Chamber Base Sample Holder Order Separately Integrated Humidty Module Figure 1 H301 OLYMPUS IX SUSP Components and Dimensions 2 Sample Holders 2 1 Available Sample Holders The following sample holders are available NOTE Please contact info oko lab com if you cannot find the sample holder you are looking for We are constantly adding new inserts to the...

Страница 4: ...X SUSP H301 SLIDING LID For easy sample loading and pipetting Included in code H301 OLYMPUS IX SUSP H301 KOEHLER LID 2 Reduces chamber height to 23 mm and allows imaging under Koehler illumination H301 INJECTION LID 2 Heated glass lid with two small openings sealed with flexible plastic allowing injec tion or permanent access to the sample Compatible only with sample holder 1x35 H301 LASER INTERLO...

Страница 5: ...r with 4 screws Screws location is indicated by letter A in Figure 4 image 2 and 3 Keep Sliding Lid fully closed to access screws labeled A in image 2 Open Sliding Lid to access screws labeled A in image 3 Distance between focal plane and upper side of the glass lid Chamber top view Sliding Lid Figure 3 H301 OLYMPUS IX SUSP Sliding Lid comes as standard 1 2 3 Sliding Lid A A Screws M2x6 Figure 4 A...

Страница 6: ...ER LID 2 Figure 5 H301 OLYMPUS IX SUSP H301 KOEHLER LID 2 3 3 Injection Lid The Injection Lid is a heated glass lid with two small openings sealed with flexible plastic allowing injection or perma nent access to the sample NOTE Compatible with sample holder 1x35 ONLY The Injection Lid is placed onto the chamber NO SCREWS or TOOLS are necessary Figure 6 shows chamber dimensions with the Injection L...

Страница 7: ...ass lid with a safety switch connecting to the laser controller It automatically turns the laser off when the lid is lifted NOTE Compatible with sample holder 1x35 ONLY The Laser Interlock Lid is placed onto the chamber NO SCREWS or TOOLS are necessary Figure 7 shows chamber dimensions with the Laser Interlock Lid Distance between focal plane and upper side of the glass lid Chamber top view H301 L...

Страница 8: ... Temperature Sensor dedicated opening Figure 8 Insertion of the temperature sensor inside the chamber 5 Insertion of Sample Holder into Chamber Sample holders fit into the chamber base and are held in place by magnets embedded within both chamber and hold er To introduce a sample holder with the proper orientation match the red dot on the holder to the one on the chamber base as illustrated in Fig...

Страница 9: ...d 60 mm petri dishes Included in the corresponding sample holder Figure 10 Magnetic locks for 35 and 60 mm dish NOTE Magnetic locks are included with sample holder Spacing rings with variable diameter are also included in according to the needs of the user Figure 11 shows how to position the ring inside the sample holder Ring selection guide diagrams rings for 35 and 60 mm dish Figure 12 and Figur...

Страница 10: ...con 35x10 Figure 12 Ring selection scheme for 35 mm dish MatTek P50G X 30 X Grainer Petri 60x15 Willco60x15 BD Falcon 60x15 Ibidi μ dish 50mm low Corning 60x15mm Figure 13 Ring selection scheme for 60 mm dish 7 Working with 1x3 and 1x2 chamber slides magnetic locks Magnetic locks prevent movement of 1 x 3 and 1 x 2 chamber slides inside of the sample holder ...

Страница 11: ... oko lab com Page 11 NOTE Magnetic locks are included with sample holder 2 1 Figure 14 Magnetic lock for 1 x 3 chamber slide 1 3 2 Figure 15 Magnetic lock for 1 x 2 chamber slide To lock the slide push simultaneously the buttons indicated with 3 ...

Страница 12: ...ercentage inside the chamber Fill the humidity module inside the chamber through the water input brass opening located on a side of the H301 OLYMPUS IX SUSP See Figure 16 using a silicon tubing 2 mm ID and a syringe NOTE Do not exceed 0 05 l Gas Input Water Input Figure 16 Connection with gas supply 9 Working with Perfusion H301 OLYMPUS IX SUSP features 8 perfusion holes for the insertion of perfu...

Страница 13: ...ect the chamber with XY stage 1 Place the chamber on the stage and tighten 4 captive screws See Figure Image 1 and Image 2 Captive screws housings are indicated with letter B in Figure Image 2 2 Use a 3mm metric Allen Wrench to tighten the four captive screws while keeping the chamber sliding lid open See Figure Image 2 Captive Screws A B 1 2 3 Figure 18 Connection of the Chamber with XY Stage ...

Страница 14: ...he chamber cools down To clean the body and the glass lid of the chamber wipe with a soft micro fiber cloth For stubborn smudges you can damp the soft micro fiber cloth with ethyl alcohol product code UN1170 Do not put any liquid di rectly on the chamber While cleaning the glass lid do not apply strong force to the surface of the glass lid because it can be damaged H301 OLYMPUS IX SUSP_User Manual...

Отзывы: