Insertion of the Sample Feedback Temperature Sensor
Insert the Sample Feedback Temperature Sensor through the dedicated opening located in the H201-UP main body
(see Figure 3, Frontal and 3D views).
1.Frontal view
2.3D view
Temperature Sensor dedicated opening
Temperature Sensor dedicated opening
Figure 3. Insertion of the temperature sensor inside the chamber.
4.
Insertion of Sample Holder into Chamber
Sample holders fit into the chamber base and are held in place by magnets embedded within both on the chamber and
on the holder. To introduce a sample holder with the proper orientation, match the red dot on the holder to the one
on the chamber base, as illustrated in Figure 4.
Insert the adapter by
superimposing the red dots
Figure 4. Introduction of the Specimen Holder inside the Chamber Base.
If you want to change the sample holder follow the steps shown in the images of Figure 5 and listed below:
1.
Open the lock spring, as shown in Figure 5 Image 2
2.
Open the chamber lid as shown in Figure 5 Image 3
3.
Remove the chamber lid by lifting it from the workspace, as shown in Figure 5 Image 4