background image

1

This  document  provides  instruction  for  the  prosthetist  in 
the installation and use of the system. A separate document, 
LimbLogic VS  Patient  Information,  is  included  with  each 
system and should be provided to the patient upon delivery 
of a prosthesis containing the LimbLogic VS.

The Vacuum Pump and fob included in this package are 
electronically keyed together and can be used only with 
each other. 

LimbLogic VS - 1600 Series is now waterproof.

Prosthetist’s 

Guide

Содержание LimbLogic VS - 1600 Series

Страница 1: ...S Patient Information is included with each system and should be provided to the patient upon delivery of a prosthesis containing the LimbLogic VS The Vacuum Pump and fob included in this package are...

Страница 2: ...as the only means of attachment of prosthesis to patient All LLVS components have passed safety testing for use as medical devices Radio enabled devices transmit with a power of less that one milliwat...

Страница 3: ...on the Vacuum Pump 11 Turning on the Fob 12 Information Screen 13 Active Mode 13 For Bilateral Applications 14 Standby Mode 14 Configuration Screen 15 Detecting a Leak 16 If the Sealing Sleeve is Punc...

Страница 4: ...oint press up and down buttons to adjust and center button to confirm Vacuum Pump battery charge level 4 To enter Standby Mode Press center button until appears Standby Mode Press center button until...

Страница 5: ...he Vacuum Pump Battery on page 9 3 Properly secure the fully charged Vacuum Pump to the socket Vacuum Plate LLV 01043 or Vacuum Pyramid LLV 01044 with M6 flathead screws Tighten the screws to 9 ft lbs...

Страница 6: ...iew Figure 1 Folding over the top edge of the liner Socket Alpha Flex Sleeve Fabric side of Alpha Liner Gel side of Alpha Liner BK Application The patient should don an Alpha Liner and Alpha Flex Slee...

Страница 7: ...iner and Alpha AK Sleeve as shown below Figure 3 Top edge of the liner is folded over the brim of the socket Socket Gel side of Alpha Liner Gel side of Alpha Liner Fabric side of AK Sleeve Figure 4 Sl...

Страница 8: ...value the Upper Set Point The difference between the Lower Set Point and the Upper Set Point is called the Vacuum Control Range The Upper Set Point cannot be set higher than the User Maximum The User...

Страница 9: ...lug will damage the plug will possibly damage the vacuum system and will void the warranty 3 Plug the charger into a power outlet The blue light on the two pronged adapter will light followed by the g...

Страница 10: ...should last for the life of the product Charging the Vacuum Pump Battery in a Vehicle To use a vehicle s power to charge the Vacuum Pump you must purchase a Power Inverter rated at 75 Watts or less t...

Страница 11: ...the pump has just been unplugged from the charger the button does not need to be pressed When you are ready for the pump to begin regulating the vacuum press the on off button again or turn on the fo...

Страница 12: ...left button Overview down button Press and hold the center button on the fob until the LimbLogic VS logo appears and the display has reached the desired brightness Release the button If the Vacuum Pu...

Страница 13: ...displayed to the left of the battery symbol as shown below the system is in Active Mode In this mode the Vacuum Pump activates as needed to maintain the vacuum settings You or the patient may use the...

Страница 14: ...RIGHT pump B Standby Mode Allows you to turn off the pump temporarily to prevent the pump from regulating the vacuum To enter Standby Mode press and hold the center button until the hourglass symbol a...

Страница 15: ...adjust the User Maximum The User Maximum cannot be set higher than 20 in hg Press the up or down button to scroll to the next screen Use the right and left buttons to adjust the Vacuum Control Range T...

Страница 16: ...tton on the fob until the hourglass symbol appears or Press the on off button on the Vacuum Pump one time If the Sealing Sleeve is Punctured The latex sleeves LLS 3N LLS 4N or LLS 5N in the Diagnostic...

Страница 17: ...cuum Pump battery has less than 15 of its capacity remaining the Vacuum Pump enters a stutter mode Stutter mode is when the pump quickly turns on and off when pulling vacuum Charge as soon as possible...

Страница 18: ...etone or similar chemicals to clean the system Please demonstrate this procedure for the patient before sending the patient home with the system Exposure to Water The Vacuum Pump is waterproof in fres...

Страница 19: ...f a cosmetic cover The kit also includes a hose barb for routing the exhaust tube into the foot shell Vacuum Plate Pyramid The Vacuum Plate LLV 01043 and Vacuum Pyramid LLV 01044 can be used for attac...

Страница 20: ...mp will not vibrate if it is already holding a vacuum when it is started up Do not press the on off button too quickly Press the button for one second then release for one second before pressing again...

Страница 21: ...nition of the button activation Wait one minute for the fob screen to turn off Press the fob center button again The fob may be too close or too far from the pump Place the fob about 2 feet away from...

Страница 22: ...tact Ohio Willow Wood With the charger plugged into an AC outlet and the pump press and hold the reset button on the charger node for 5 seconds then release If the green LED does not blink or turn sol...

Страница 23: ...nother room The configurable prong may not be seated properly in the back of the charger s two pronged adapter Unplug the charger from the AC outlet Remove the charger s prong using the sliding latch...

Страница 24: ...l the hole with epoxy then drill a new 9 32 hole in the center The leak was coming from the Vacuum Pump Contact Ohio Willow Wood If the leak does NOT go away If the leak goes away If the leak goes awa...

Страница 25: ...Wood s recommendations and will void the two year warranty Extremely high and abusive activities are defined as activities such as skydiving karate and judo activities that could result in injury to...

Страница 26: ...al property rights reflected or incorporated in its physical products regardless of the transfer of the physical products to another party or parties There are no field serviceable parts inside the Va...

Страница 27: ...his product complies with IEC 60601 1 1998 A1 1991 A2 1995 EN 300 440 1 V1 3 1 2001EMC EN 301 489 1 V1 7 1 2006 EN 301 489 3 V1 4 1 2002 EMC and EN55022 1998 A1 2000 A2 2003 IEC 60601 1 Classification...

Страница 28: ...ically tested for wound healing or other uses Ohio Willow Wood does not currently support the use of the LLVS for uses other than suspension Use only the bolts supplied by The Ohio Willow Wood Company...

Страница 29: ...m Pumps use Ohio Willow Wood charger part number LLV 01010 Use with other products may result in harm to the device the user or both and void the warranty Refer to product manuals to verify compatibil...

Страница 30: ...f the FCC Rules and RSS 210 Annex 2 9 This module has been designed for use by the Ohio Willow Wood Co This module shall only be designed into products by The Ohio Willow Wood Co Warning Changes or mo...

Страница 31: ...12V DC Input current 37 mA Operating temperature range 20 C to 50 C Frequency range 2400 2483 MHz Antenna Permanently soldered to Module Transmit power 2mW Communication type Frequency Hopping Spread...

Страница 32: ...etic environments specified below The customer or the user of the LLVS should assure it is used in such an environment Emissions Test Compliance Guidance RF Emissions Group 2 The LLVS must emit electr...

Страница 33: ...kV Differential Mains power quality IEC 61000 4 5 Mode Mode should be that of a typical 2 kV Common 2 kV Common commercial or hospital Mode Mode environment Voltage Dips 5 UT 5 UT Mains power quality...

Страница 34: ...Hz to 2 5 GHz IEC 61000 4 3 Where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer and d is the recommended separation distance in meters m...

Страница 35: ...mated using the equation applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer NOTE 1 At 80...

Страница 36: ...s Pending OHIO WILLOW WOOD free the body free the spirit 15441 Scioto Darby Road Mt Sterling OH 43143 phone 740 869 3377 800 848 4930 fax 740 869 4374 www owwco com Ohio Willow Wood Company B V Keizer...