8 550102
(Rev.
12)
10/2021
4. If the vacuum port tubing was contaminated, open the unit and inspect the following
parts for contamination (see Maintenance section below for opening the unit):
All Internal Tubing: If contaminated, replace.
Regulator Assembly: If contaminated, clean or replace according to procedure below.
Solenoid Manifold Assembly If contaminated, clean or replace according to procedure on
page 9.
Pump Assembly: If contaminated, clean or replace according to procedure on page 10.
9. Maintenance
Tools required to service the Instavac:
Pliers
Phillips screwdriver 11/32” nut driver
5/16” nut driver
5/64” Allen wrench Adjustable wrench
1/16” Allen wrench
Flathead screwdriver
To access the internal components of the Instavac, remove the two (2) Phillips head
screws located in the rear of the unit near the back assembly. Tilt the cover forward.
CAUTION: NEVER DISASSEMBLE THE INSTAVAC
WHEN THE POWER CORD IS
CONNECTED TO AN ELECTRICAL OUTLET.
NOTE
: After performing any service or repair of the Instavac, including preventative
maintenance and cleaning for overflow, verify that the device performance specifications
are met prior to returning to service. Follow steps in section 10 to verify and adjust the
device settings as necessary.
PREVENTATIVE MAINTENANCE
The Instavac is warranted for its intended life of three (3) years. Performing periodic
preventative maintenance extends the life of the Instavac portable aspirator. Preventative
maintenance is required once per year regardless of frequency of use and is recommended
every six months with frequent use.
A Preventative Maintenance Kit is available (Part Number AI5801ASY). It contains all
necessary items and instructions for performing the preventative maintenance procedure.
VACUUM REGULATOR REPLACEMENT
1. Tilt the shroud assembly forward off of the Instavac
base (make a note of the
orientation of all tubing and the regulator body).
2. Disconnect all three (3) tubing connections from the regulator body.
3. Using a small flathead screwdriver to loosen the set screw on the vacuum regulator
knob, unscrew and remove the vacuum regulator knob from the control panel.
4. Remove the outside locknut that secures the regulator to the control panel. The
regulator can now be removed.
5. To mount the new regulator, remove the knob from the body by loosening the set
screw on the Vacuum regulator knob (use a flathead screwdriver).
6. Mount the new regulator body to the control panel. Make sure the fittings are in the
(Intermittent models only):
Содержание 756000
Страница 1: ...Operator Maintenance Manual 550102 Rev 12 10 2021 Multipurpose Suction Unit Medical Equipment...
Страница 16: ...16 550102 Rev 12 10 2021 Figure 2 Instavac I 756000 756010 See page 18 for Parts List...
Страница 17: ...550102 Rev 12 10 2021 17 Figure 3 Instavac I 756000 756010 See page 18 for Parts List...
Страница 19: ...550102 Rev 12 10 2021 19 Figure 4 Instavac C 757000 757010 See page 22 for Parts List...
Страница 20: ...20 550102 Rev 12 10 2021 Figure 5 Instavac C 757000 757010 See page 22 for Parts List...
Страница 21: ...550102 Rev 12 10 2021 21 Figure 6 Instavac C 757000 757010 See page 22 for Parts List...
Страница 23: ...550102 Rev 12 10 2021 23 Notes...