EN-8 OHAUS Valor
TM
1000W Series
4.3 Restore Settings to Default
Press and hold SET, then
h
until [SEt-F2] is displayed. Press YES to enter and view
default user settings [F1-0 x].
Press YES to confirm displayed setting and proceed to next menu item.
Press SET to save and exit.
4.4 Set Tare Capacity
Press and hold SET, then
h
until [SEt-F3] is displayed. Press YES to enter and view
tare capacity settings [tArE x], where x:
0 = 2/3 of scale capacity
1 = 100% of scale capacity
Press
ji
to toggle through available settings, then press SET to save and exit.
5. SCALE SETUP AND SPAN CALIBRATION
For best results, calibrate the scale at regular intervals. Temperature changes,
gravity variations and abuse are some reasons why a scale may need recalibration.
Enter calibration as follows:
•
Turn the scale on.
•
During the start-up display [000000…999999], press and hold SET until
[FS--SP] is displayed.
•
Press YES to display [P x], where x = position of decimal point
•
Press
ji
to toggle through settings: 0, 1, 2, 3, 4
•
Press YES to select, then display [d x], where x = scale divisions
•
Press
ji
to toggle through settings: 1, 2, 5, 10, 20, 50
•
Press YES to select, then display scale capacity [xxxxxx] (kg)
•
Press YES to select, then display [noLoAd]
•
Make sure pan is empty, then press YES to display [LoAd]
•
Press YES to select, then display default or last used span weight [xxxxxx] (kg)
•
Press
h
,
j
,
i
to edit the displayed value (as needed)
•
Place indicated weight on the scale, then press YES
•
[CALEnd] is momentarily displayed to indicate calibration was successful, then
exit to weighing mode
•
To abort calibration at any point, turn the scale off.
Examples:
Parameters:
3 x 0.0005 kg
6 x 0.001 kg
15 x 0.002 kg
30 x 0.005 kg
30 x 0.01 kg
P
4 3 3 3 3
d
5 1 2 5 10
Capacity
3.000 6.000 15.000 30.000
30.000