Valor 5000 Series
EN-22
4.4 Service Information
If the Troubleshooting section does not resolve or describe your problem, contact your authorized
Ohaus Service Agent. For Service assistance in the United States and Canada, please call Ohaus
Corporation toll free at (800) 526-0659 from 8 AM to 5 PM Eastern Time. An Ohaus Product Service
Specialist will be available to help you.
4.5 Accessories
Description
Part Number
Foot Switch
71169633
TABLE 4-2. ERROR CODES.
Error Code
Cause
Remedy
Err 2 (Underload)
Missing pan.
Install the pan and press the zero
button.
Err 9 (Checksum error)
Err14 (Zero error)
Err 21 (Checksum error)
CAL E (Calibration error)
Configuration data
checksum error.
Service required.
Zero limit exceeded
during power up.
Remove load from pan.
Calibration data
checksum error.
Check menu settings and recalibrate.
If error persists, service is required.
Wrong calibration weight
values used.
Calibrate using correct calibration
weights.
Err 1 (Overload)
Load exceeds capacity.
Reduce load.