Valor 3000 Series Scales EN-5
3.2 Percent Mode
Use this mode to measure the weight of a sample as a percentage of a reference
weight.
1. Press
MODE/UNIT/Menu
until Set.rEF and the % icon are displayed.
2. Press
No
to use the reference weight in memory and proceed to step 6.
3. Press
Yes
to establish a new reference. Scale display will now blink until
reference is stored.
4. Put an empty container on the pan, then press
O/T/On/Off
to tare the container
weight.
5. Add the desired amount of material to the container, then press
ENTER/Cal
to
store the displayed weight. The display shows 100%.
6. Replace the reference material with new material. The display shows the
percentage of the sample compared to reference weight.
3.3 Specific Gravity Mode
Use this mode to measure the specific gravity of a sample.
1. Press
MODE/UNIT/Menu
until Set.rEF and the S icon are displayed.
2. Press
No
to use the reference weight in memory and proceed to step 6.
3. Press
Yes
to establish a new reference weight. Scale display will now blink until
reference is stored.
4. Put an empty container on the pan, then press
O/T/On/Off
to tare the
container
weight.
5. Fill the container with water, then press
ENTER/Cal
to store the displayed
weight. The display shows the water weight with the S icon blinking.
6. Replace the water with the same volume of sample material, then press
ENTER/Cal
. The display shows the specific gravity of the sample.
3.4 Fluid Ounce Mode
Use this mode to measure the fluid ounce volume of a sample.
1. Press
MODE/UNIT/Menu
until the FL icon is displayed.
2. Put an empty container on the pan, then press
O/T/On/Off
to tare the
container
weight.
3. Fill the container with sample material. The display shows the volume of the
sample.