EN-14 T21P Indicator
6. SERIAL COMMUNICATION
The Indicator includes an RS232 serial communication interface. The setup of
RS232 operating parameters are more fully explained in Section 4.3 [
F1-6
] and
[
F1-7
]. The physical hardware connection is explained in Section 2.2.2.
The interface enables display data to be sent to a computer or printer.
6.1 OUTPUT FORMAT
The default serial output format is shown below.
Field: Space Polarity Weight
Unit
Stability
Legend
CR
LF
Length: 2
1 7 2 1
1 1
1
Example:
- 3 8 . 5 6 0 k g / G
Definitions:
Polarity, “-“ sign if negative, blank if positive.
Weight, up to 6 numbers and 1 decimal, right justified, leading zero blanking.
Units, up to 2 characters.
Stability, “/” character is printed if not stable, blank if stable.
Legend, up to 3 characters: G= gross weight, N= net weight, T= tare.