EN-4
STARTER 300C Portable Conductivity Meter
2 INSTALLATION
Carefully unpack the meter.
2.1
Package contents
The model ST300C-B (Basic Model)should have the following items
In addition to ST300C-B content, the ST300C Kit also includes the conductivity electrode
STCON3:
The best measurement range for STCON3 4-ring (4-pole) cond. electrode is 70
μ
S/cm-200mS/cm.
In this range the measurement accuracy can reach 0.5%.
If you use STCON3 to measure pure water in which conductivity value is below 70
μ
S/cm, the
measurement will be less accurate (approximately 1% to 5%).
Does not use STCON3 to measure conductivity value less than 2
μ
S/cm, the accuracy is > +/-5%.
Accessories:
Model
Description
P/N
STCON3
4-ring cond. electrode
(70
μ
S/cm-200mS/cm, 0.5% accuracy; )
( 2-70
μ
S/cm, 1-5% accuracy)
83033972
STCON7
Pure water cond. electrode
(0.02
μ
S/cm-200
μ
S /cm; 0.02
μ
S/cm
accuracy)
30080693
Standard Conduct 10
μ
S/cm 250ml
30100441
Standard Conduct 84
μ
S/cm 250ml
30100442
Standard Conduct 1413
μ
S/cm 250ml
30100443
Standard Conduct12.88mS/cm 250ml
30100444
Portable Bag for portable meters
30031635
ST300C-B
Units
STARTER 300C meter
1
AAA battery
4
1413
μ
S/cm standard solution
1 bottle (about 20ml)
12.88 mS/cm standard solution
1 bottle (about 20ml)
Electrode Clip
1
IP54 electrode caps
1 set
Wrist Strap
1
STCON3
1
www.calcert.com
1.888.610.7664
0
5
10
15
20
25
30