EN-8 STARTER 300 Portable pH Meter
3
STARTER 300 SETUP
3.1
Set temperature unit and MTC value
Please note:
If a temperature electrode is used, Automatic Temperature Compensation (
ATC)
and the sample
temperature are displayed on the screen. You may then choose to skip MTC setup (below).
If the meter does not detect a temperature electrode or one is not used, the meter automatically
switches to Manual Temperature Compensation (
MTC
) mode and
MTC
appears on the screen.
MTC can be set as follows:
Power the meter on by pressing
Press and hold
until the setup icon
appears on the display and the current
temperature unit blinks (
°
C or
°
F).
Press
or
to switch between
°
C and
°
F .
Press
to confirm your selection.
Then
Continue with MTC temperature setting by using
or
to adjust temperature
compensation accordingly
Press button-
to confirm the setting
Press
to return to the measurement screen.
The default MTC temperature value setting is 25 °C (77
°
F).
Note:
°
C = 5/9 (
°
F - 32)
3.2
Selecting a predefined buffer group
After confirming the MTC value (or if you are using ATC) you have the option of making a buffer
group selection. In most cases the default buffer group will be used, buffer group
b1
.
If it is necessary to switch the buffer group, use button-
or button-
to select a buffer
group among 4 buffer groups shown in section 4.1.1 in this manual. Press button-
to confirm
the setting or press button-
to return to the measurement screen.
The default buffer group is
b1
.
b1
1.68
4.01
7.00
10.01