EN-16 SF40A Printer
Problem / Message
Possible cause
Correction
--------
Not Executable
------
Instrument cannot
execute printer
instruction.
Check the balance.
----
Date/Time Not Set
---
---------
Memory Error
---------
Printer error
messages.
Check the instrument.
Reset the printer.
Set date/time
Back-up battery
depleted
Please contact OHAUS
service.
---------
Memory full
----------
Statistics or
Totaling Memory is
full.
Start a new Statistics or
Totaling application.
--
Load/Unload Weight
---
No sample or the
same sample is on
the weighing pan.
Put a new sample or
remove it and put it
again on the pan.
Unit change not
allowed!
Changed unit
during Statistics or
Totaling
application.
Reset the balance.
Change the unit back to
the first one.
---------
Out of range
---------
Sample weight is
out of range
Sample weight must be
in the range of 70% -
130% of the current
average value.
DOMINIQUE DUTSCHER SAS
Содержание SF40A
Страница 2: ...D O M I N I Q U E D U T S C H E R S A S...
Страница 13: ...EN 10 SF40A Printer 4 Menu D O M I N I Q U E D U T S C H E R S A S...
Страница 23: ...EN 20 SF40A Printer D O M I N I Q U E D U T S C H E R S A S...
Страница 34: ...SF40A Drucker DE 31 4 Men D O M I N I Q U E D U T S C H E R S A S...
Страница 44: ...SF40A Drucker DE 41 D O M I N I Q U E D U T S C H E R S A S...
Страница 56: ...SF40A Impresora ES 75 4 Me n D O M I N I Q U E D U T S C H E R S A S...
Страница 66: ...SF40A Impresora ES 85 D O M I N I Q U E D U T S C H E R S A S...
Страница 78: ...SF40A Imprimante FR 53 4 MENU D O M I N I Q U E D U T S C H E R S A S...
Страница 88: ...SF40A Imprimante FR 63 D O M I N I Q U E D U T S C H E R S A S...
Страница 100: ...SF40A Stampante IT 97 4 Menu D O M I N I Q U E D U T S C H E R S A S...
Страница 111: ...D O M I N I Q U E D U T S C H E R S A S...